DOI 10.1583/09-2987mr.1

Медицинская библиотека · Journal of endovascular therapy : an official journal of the International Society of Endovascular Specialists · 2010

Feasibility and limitations of an automated 2D-3D rigid image registration system for complex endovascular aortic procedures.

Carrell Tom W G, Modarai Bijan, Brown James R I, Penney Graeme P

Оценочное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

PURPOSE: To examine the feasibility of an automated 2-dimensional (2D) to 3- dimensional (3D) image registration system to simplify the navigational challenges faced in complex endovascular aortic procedures.

METHODS

An automated 2D-3D image registration system was used to overlay pre-acquired 3D computed tomography images onto fluoroscopy images taken during endovascular aneurysm repair. Errors between the 3D overlay and digital subtraction angiograms were measured and correlated with aortic neck angulation. A mean discrepancy < or =3 mm was considered clinically acceptable.

RESULTS

There was a strong correlation between maximum neck angulation and maximum registration error (Pearson's r = 0.75). Aortas with a maximum neck angulation < or =30 degrees had a mean error of 2.5+/-1.2 mm, whereas aortas with neck angulation >30 degrees had a mean error of 6.2+/-2.5 mm (p<0.0001).

CONCLUSION

The major source of registration errors is aortic deformation caused by the presence of the introducer and endovascular graft. Further work is required if this technology is to be routinely applied to severely angulated aortic anatomy.

Для цитирования: Carrell Tom W G, Modarai Bijan, Brown James R I, Penney Graeme P. Feasibility and limitations of an automated 2D-3D rigid image registration system for complex endovascular aortic procedures.. Journal of endovascular therapy : an official journal of the International Society of Endovascular Specialists. 2010. DOI: 10.1583/09-2987mr.1.