DOI 10.1097/rhu.0000000000002372
The Usefulness of Antinuclear Antibody Multiplex Immunoassay for Screening of Rheumatological Diseases: A Real-World Population-Based Study.
Аннотация
OBJECTIVE
To evaluate the diagnostic performance of BioPlex 2200 multiplex immunoassay (BI) for detecting antinuclear antibodies (ANA), compared with the gold standard, the indirect immunofluorescence assay (IIF), as a screening methodology for autoimmune diseases within a general population.
METHODS
Cross-sectional, database-based analysis of the Clalit Health Services nationwide registry. Data were extracted from the Clalit Health Services database between 2014 and 2023. We included adult outpatients without a known rheumatic disease who underwent both IIF and BI testing simultaneously, ordered for nonspecific signs and symptoms primarily by primary-care physicians.
RESULTS
Among the 391,366 individuals tested, 67% were female, and the median age at testing was 49 years. Annual ANA testing volume increased from 23,802 in 2014 to 76,898 in 2023 (3.2-fold). Of the total, 33,234 (8.4%) tests were false negatives (BI-negative while IIF-positive), 42,021 (10.7%) were false positives (BI-positive while IIF-negative), 302,439 (77.2%) were true negatives, and 13,672 (3.5%) were true positives. BI demonstrated a sensitivity of 29.1% (95% CI: 28.7-29.6) and a specificity of 87.8% (95% CI: 87.7-87.9). The positive predictive value was 24.5% and the negative predictive value 90.1%.
CONCLUSION
Our findings do not support the use of the BI multiplex immunoassay as a primary ANA screening tool in the primary-care setting. On the basis of these findings, the Clalit Health Services laboratory network discontinued simultaneous IIF and BI testing.
KEY FINDINGS: The BI showed low sensitivity despite relatively high specificity, resulting in a substantial proportion of false-negative results compared with IIF. These findings indicate that it is not suitable as a primary screening tool for ANA in the primary-care setting.