Clinical biomechanics (Bristol, Avon) · 2015

Biomechanical effects of fusion levels on the risk of proximal junctional failure and kyphosis in lumbar spinal fusion surgery.

СтатьяПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Park Won Man, Choi Dae Kyung, Kim Kyungsoo, Kim Yongjung J, Kim Yoon Hyuk

Аннотация

BACKGROUND: Spinal fusion surgery is a widely used surgical procedure for sagittal realignment. Clinical studies have reported that spinal fusion may cause proximal junctional kyphosis and failure with disc failure, vertebral fracture, and/or failure at the implant-bone interface. However, the biomechanical injury mechanisms of proximal junctional kyphosis and failure remain unclear.

METHODS: A finite element model of the thoracolumbar spine was used. Nine fusion models with pedicle screw systems implanted at the L2-L3, L3-L4, L4-L5, L5-S1, L2-L4, L3-L5, L4-S1, L2-L5, and L3-S1 levels were developed based on the respective surgical protocols. The developed models simulated flexion-extension using hybrid testing protocol.

FINDINGS: When spinal fusion was performed at more distal levels, particularly at the L5-S1 level, the following biomechanical properties increased during flexion-extension: range of motion, stress on the annulus fibrosus fibers and vertebra at the adjacent motion segment, and the magnitude of axial forces on the pedicle screw at the uppermost instrumented vertebra.

INTERPRETATIONS: The results of this study demonstrate that more distal fusion levels, particularly in spinal fusion including the L5-S1 level, lead to greater increases in the risk of proximal junctional kyphosis and failure, as evidenced by larger ranges of motion, higher stresses on fibers of the annulus fibrosus and vertebra at the adjacent segment, and higher axial forces on the screw at the uppermost instrumented vertebra in flexion-extension. Therefore, fusion levels should be carefully selected to avoid proximal junctional kyphosis and failure.