Corrosion of Modular Dual-Mobility Acetabular Components Leading to Acetabular Bone Loss and Protrusio.

Yang JaeWon, Dasari Suhas P, Chansky Howard A, Hernandez Nicholas M

Клинический случайСтатьяПеревод в очереди

Аннотация

Dual-mobility bearings are being increasingly utilized in total hip arthroplasty. Contemporary modular designs often feature inner cobalt-chromium liners that are seated in outer titanium acetabular shells. However, mating of these 2 dissimilar metals may lead to complications. We present a case report of a patient who was found to have osteolysis and acetabular protrusio due to an adverse local tissue reaction from corrosion between a cobalt-chromium liner and titanium acetabular shell that appeared to be well-seated clinically and radiographically. This case suggests that implant failure and associated adverse local tissue reactions may occur even in seemingly well-seated modular dual-mobility implants.

Для цитирования: Yang JaeWon, Dasari Suhas P, Chansky Howard A, Hernandez Nicholas M. Corrosion of Modular Dual-Mobility Acetabular Components Leading to Acetabular Bone Loss and Protrusio.. Arthroplasty today. 2024. DOI: 10.1016/j.artd.2024.101543.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.