DOI 10.1007/s10439-026-04258-8
Multidirectional Optical Bone Densitometry Using a Simulation-Based Machine Learning Model: Experimental Validation with Bone Phantoms.
Аннотация
PURPOSE: While the multidirectional optical bone densitometry approach, originally proposed based on simulation, has been previously reported, its experimental validation has not yet been demonstrated. This study addresses this gap by validating the method using bone phantoms.
METHODS
These phantoms were fabricated with Intralipos, agarose, and bovine cortical bone fragments. By varying the mixing ratio of cortical bone fragments, phantoms with different bone densities were prepared. Near-infrared laser light with a wavelength of 850 nm was directed onto the phantoms from three directions, and the resulting backward, lateral, and forward light intensity distributions on the phantom surface were recorded. Directional light intensity values were extracted from these distributions and used as features for the simulation-based machine learning model to predict bone density.
RESULTS
Linear regression between the predicted and reference bone densities yielded an r2 of 0.85.
CONCLUSION
These findings demonstrate the feasibility of the proposed method as a proof-of-concept validation and suggest a possible direction for future development of optical bone density assessment methods.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.