Isoforce: A new outrigger system for static progressive orthotic interventions of the proximal interphalangeal joint with constant force transmission-Results of a biomechanical study.
СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Marrel Marion, Jörn Good Ulla, Marks Miriam, Herren Daniel B, Goldhahn Jörg
Аннотация
STUDY DESIGN: Bench research-biomechanical study.
INTRODUCTION: Static progressive orthotic devices are efficient in treating contractures. However, current outriggers are unable to keep force transmission and the force application angle (FAA) constant.
PURPOSE OF THE STUDY: To evaluate the biomechanical performance of the Isoforce outrigger, a novel extension orthosis.
METHODS: A hand model was used to measure the required force at the outrigger and FAA, while simulating resolution of different contracture angles. We also tested feasibility in a small patient series.
RESULTS: The force required with the Isoforce device never exceeded 2.4 N, and the FAA did not change more than 6°. Corresponding figures for the reference devices exceeded 16 N and 20°. The 7 patients testing the Isoforce extension device showed an extension deficit that decreased from 40° at baseline to 25° at 6 weeks. They rated the device as very comfortable to wear.
CONCLUSIONS: Isoforce maintains constant force transmission and FAA throughout the full range of motion, promotes the lengthening of contracted structures, and is comfortable to wear.
LEVEL OF EVIDENCE: Not applicable.
INTRODUCTION: Static progressive orthotic devices are efficient in treating contractures. However, current outriggers are unable to keep force transmission and the force application angle (FAA) constant.
PURPOSE OF THE STUDY: To evaluate the biomechanical performance of the Isoforce outrigger, a novel extension orthosis.
METHODS: A hand model was used to measure the required force at the outrigger and FAA, while simulating resolution of different contracture angles. We also tested feasibility in a small patient series.
RESULTS: The force required with the Isoforce device never exceeded 2.4 N, and the FAA did not change more than 6°. Corresponding figures for the reference devices exceeded 16 N and 20°. The 7 patients testing the Isoforce extension device showed an extension deficit that decreased from 40° at baseline to 25° at 6 weeks. They rated the device as very comfortable to wear.
CONCLUSIONS: Isoforce maintains constant force transmission and FAA throughout the full range of motion, promotes the lengthening of contracted structures, and is comfortable to wear.
LEVEL OF EVIDENCE: Not applicable.