MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.wneu.2024.01.168

Медицинская библиотека · World neurosurgery · 2024

Risk Factors for Medial Breach During Robotic-Assisted Cortical Bone Trajectory Screw Insertion.

Nagata Kosei, Gum Jeffrey L, Brown Morgan, Daniels Christy, Hines Bren, Carreon Leah Y, Glassman Steven D

СтатьяПеревод в очереди

Аннотация

OBJECTIVE

We describe the incidence of, and identify the risk factors for, a medial breach of the pedicle wall during robotic-assisted cortical bone trajectory (RA-CBT) screw insertion.

METHODS

We analyzed a consecutive series of adult patients who underwent RA-CBT screw placement from January 2019 to July 2022. To assess the pedicle wall medial breach, postoperative computed tomography (CT) images were analyzed. Patient demographic data and screw data were compared between patients with and without a medial breach. The Hounsfield units (HUs) on the L1 midvertebral axial CT scan was used to evaluate bone quality.

RESULTS

Of 784 CBT screws in 145 patients, 30 (3.8%) had a medial breach in 23 patients (15.9%). One screw was grade 2, and the others were grade 1. Patients with a medial breach had a lower HU value compared with the patients without a medial breach (123.3 vs. 150.5; P = 0.027). A medial breach was more common in the right than left side (5.5% vs. 2.0%; P = 0.014). More than one half of the screws with a medial breach were found in the upper instrumented vertebra (UIV) compared with the middle construct or lowest instrumented vertebra (6.7% vs. 1.3% vs. 2.7%; P = 0.003). Binary logistic regression showed that low HU values, right-sided screw placement, and UIV were associated with a medial breach. No patients returned to the operating room for screw malposition. No differences were found in the clinical outcomes between patients with and without a medial breach.

CONCLUSIONS

The incidence of pedicle wall medial breach was 3.8% of RA-CBT screws in the postoperative CT images. A low HU value measured in the L1 axial image, right-sided screw placement, and UIV were associated with an increased risk of medial breach for RA-CBT screw placement.

Для цитирования: Nagata Kosei, Gum Jeffrey L, Brown Morgan, Daniels Christy, Hines Bren, Carreon Leah Y, Glassman Steven D. Risk Factors for Medial Breach During Robotic-Assisted Cortical Bone Trajectory Screw Insertion.. World neurosurgery. 2024. DOI: 10.1016/j.wneu.2024.01.168.