Archives of orthopaedic and trauma surgery · 2024
Performance of medial pivot, posterior stabilized and rotating platform total knee arthroplasty based on anteroposterior stability and patient-reported outcome measures; a multicentre double-blinded randomized controlled trial of 210 knees.
СтатьяРандомизированное исследованиеМногоцентровое исследованиеНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
De Groot J D, Brokelman R B G, Lammers P G, Van Stralen G M J, Kooijman C M, Hokwerda S T
Аннотация
BACKGROUND: Despite advancements in total knee arthroplasty (TKA), 10-20% of patients remain dissatisfied after surgery. Improved anteroposterior (AP) stability provided by medial pivot (MP) implants may theoretically lead to higher patient satisfaction.
METHODS: AP stability and patient-reported outcome measures (PROMs) at one-year postsurgery were compared between patients who underwent TKA with MP- (n = 121), posterior stabilized (PS; n = 53) and rotating platform (RP; n = 57) implants in a double-blind multicentre randomized controlled trial (Dutch Trial Register: NL6856, 21-02-2018). AP stability was assessed at 30°, 60° and 90° of knee flexion using a KT-2000 arthrometer. PROMs were measured preoperative and one-year postsurgery.
RESULTS: MP-TKA provided significant better AP stability at early flexion (30°) compared to PS- and RP-TKA (median [IQR]; 1.79 [1.14-2.77] mm vs. 3.31 [2.51-4.08] mm vs. 2.82 [1.80-4.03] mm, p < 0.001). Additionally, MP-TKA provided significant better AP stability at mid-flexion (60°) compared to PS-TKA (1.75 [1.23-2.36] mm vs. 2.14 [1.49-2.83] mm, p = 0.014). PROM improvements were comparable between implant designs. AP laxity of ≥ 4 mm at early flexion was independently of implant design associated with significantly worse Kujala scores. The incidence of ≥ 4 mm AP laxity at any knee angle was however not significantly different between implant designs.
CONCLUSION: MP-, PS- and RP-TKA all provide excellent and comparable results. Although MP-TKA provided better AP stability at early flexion compared to PS- and RP-TKA, this was found to be unrelated to improved PROMs in favour of MP-TKA. More studies focusing on early and mid-flexion performance based differences between MP and other TKA designs are required to confirm our findings. Other non-implant related factors may play a more important role in the performance of TKA and are potentially worthwhile examining.
METHODS: AP stability and patient-reported outcome measures (PROMs) at one-year postsurgery were compared between patients who underwent TKA with MP- (n = 121), posterior stabilized (PS; n = 53) and rotating platform (RP; n = 57) implants in a double-blind multicentre randomized controlled trial (Dutch Trial Register: NL6856, 21-02-2018). AP stability was assessed at 30°, 60° and 90° of knee flexion using a KT-2000 arthrometer. PROMs were measured preoperative and one-year postsurgery.
RESULTS: MP-TKA provided significant better AP stability at early flexion (30°) compared to PS- and RP-TKA (median [IQR]; 1.79 [1.14-2.77] mm vs. 3.31 [2.51-4.08] mm vs. 2.82 [1.80-4.03] mm, p < 0.001). Additionally, MP-TKA provided significant better AP stability at mid-flexion (60°) compared to PS-TKA (1.75 [1.23-2.36] mm vs. 2.14 [1.49-2.83] mm, p = 0.014). PROM improvements were comparable between implant designs. AP laxity of ≥ 4 mm at early flexion was independently of implant design associated with significantly worse Kujala scores. The incidence of ≥ 4 mm AP laxity at any knee angle was however not significantly different between implant designs.
CONCLUSION: MP-, PS- and RP-TKA all provide excellent and comparable results. Although MP-TKA provided better AP stability at early flexion compared to PS- and RP-TKA, this was found to be unrelated to improved PROMs in favour of MP-TKA. More studies focusing on early and mid-flexion performance based differences between MP and other TKA designs are required to confirm our findings. Other non-implant related factors may play a more important role in the performance of TKA and are potentially worthwhile examining.