DOI 10.1016/j.clinbiomech.2026.106769

Медицинская библиотека · Clinical biomechanics (Bristol, Avon) · 2026

Threshold screw insertion torque for carbon fibre-reinforced polyetheretherketone and titanium (Ti-6Al-4V) locking plate constructs.

Nasr Ramzi, Pegg Elise, Zaribaf Fedra, Fletcher James

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Locking plate constructs are integral to modern fracture fixation, particularly in osteoporotic bone. Carbon-fibre reinforced polyetheretherketone (CFR-PEEK) plates offer a stiffness profile closer to cortical bone than titanium, potentially improving the mechanical environment for healing. However, the optimal insertion torque for locking screws in CFR-PEEK plates is unclear. This study investigated the influence of insertion torque on construct performance in CFR-PEEK plates compared to titanium alloy (Ti-6Al-4V) plates.

METHODS

Locking screws (3.5 mm, Ti-6Al-4V) were inserted into Ti-6Al-4V and CFR-PEEK plates at six torque levels (0.5-3.0 Nm). Construct strength was assessed via axial push-out and cantilever bending tests. Video analysis was used to correlate insertion torque with screw rotation. Additional testing was performed on two-screw CFR-PEEK constructs at three torque levels (0.5, 1.5, 2.5 Nm).

FINDINGS: Insertion torque was linearly related to screw rotation within the tested range (0.5-3.0 Nm). In CFR-PEEK single screw constructs, higher torque (≥2 Nm) improved push-out strength (p < 0.05), while 1.5 Nm yielded the highest cantilever strength (p < 0.05). Ti-6Al-4V constructs showed a similar trend. In two-screw CFR-PEEK constructs, no significant differences were found in performance across torques for either push-out or cantilever testing (p > 0.5).

INTERPRETATION: Beyond an initial threshold (∼0.5 Nm), increased insertion torque did not consistently enhance construct performance. These findings indicate that moderate torques (∼1.5 Nm) optimise performance while reducing the risk of implant damage, support the mechanical reliability of CFR-PEEK plates.

Для цитирования: Nasr Ramzi, Pegg Elise, Zaribaf Fedra, Fletcher James. Threshold screw insertion torque for carbon fibre-reinforced polyetheretherketone and titanium (Ti-6Al-4V) locking plate constructs.. Clinical biomechanics (Bristol, Avon). 2026. DOI: 10.1016/j.clinbiomech.2026.106769.