Biomechanical effects of posterior pedicle screw-based instrumentation using titanium versus carbon fiber reinforced PEEK in an osteoporotic spine human cadaver model.

Oikonomidis Stavros, Greven Johannes, Bredow Jan, Eh Madita, Prescher Andreas, Fischer Horst, Thüring Johannes, Eysel Peer, Hildebrand Frank, Kobbe Philipp, Scheyerer Max Joseph, Herren Christian

Сравнительное исследованиеСтатьяПеревод в очереди

Аннотация

BACKGROUND

Aim of this biomechanical investigation was to compare the biomechanical effects of a carbon fiber reinforced PEEK and titanium pedicle screw/rod device in osteoporotic human cadaveric spine.

METHODS

Ten human fresh-frozen cadaveric lumbar spines (L1-L5) have been used and were randomized into two groups according to the bone mineral density. A monosegmental posterior instrumentation (L3-L4) using titanium pedicle screws and rods was carried out in group A and using carbon fiber reinforced PEEK in group B. A cyclic loading test was performed at a frequency of 3 Hz, starting with a peak of 500 N for the first 2000 cycles, up to 950 N for 100,000 cycles under a general preload with 100 N. All specimens were evaluated with regard to a potential collapse of the implanted pedicle screws. A CT supported digital measurement of cavities around the pedicle at 3 defined measuring points was performed. Finally, the maximum zero-time failure load of all specimens was determined using a universal testing machine (80% Fmax).

FINDINGS: Regarding maximum axial force (group A: 2835 N, group B: 3006 N, p = 0.595) and maximum compression (group A: 11.67 mm, group B: 15.15 mm, p = 0.174) no statistical difference could be shown between the two groups. However, significant smaller cavity formation around the pedicle screws could be observed in group B (p = 0.007), especially around the screw tip (p < 0.001).

INTERPRETATION: Carbon fiber reinforced PEEK devices seem to be advantageous in terms of microscopic screw loosening compared to titanium devices.

Для цитирования: Oikonomidis Stavros, Greven Johannes, Bredow Jan, Eh Madita, Prescher Andreas, Fischer Horst, Thüring Johannes, Eysel Peer, Hildebrand Frank, Kobbe Philipp, Scheyerer Max Joseph, Herren Christian. Biomechanical effects of posterior pedicle screw-based instrumentation using titanium versus carbon fiber reinforced PEEK in an osteoporotic spine human cadaver model.. Clinical biomechanics (Bristol, Avon). 2020. DOI: 10.1016/j.clinbiomech.2020.105153.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.