MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1055/s-2008-1051438

Медицинская библиотека · Zeitschrift fur Orthopadie und ihre Grenzgebiete · 1981

[Wear measurements of hip prostheses with UHMW polyethylene (RCH-1000) socket and chromium plated protasul-10 head (author's transl)].

Afifi K F, Jacob H A

Аннотация на английскомСтатьяПеревод в очереди

Аннотация

An estimation of wear was carried out on 100 of 340 UHMW polyethylene sockets (RCH 1000) that were implanted with Protasul-10 prostheses furnished with a chromium plated head. This was effected by measuring the position of the head relative to the socket as seen on radiographs during a follow-up period of 4 years. The method employed was verified in 3 cases (in which the socket had been removed) by measurements on silastic casts of the bearing surface. Practically no difference in the results was observed. The average wear rate amounted to 0.16 mm/year over an observation period of 4 years. Furthermore, it has been observed that para-articular calcification seems to be independent of the amount of abrasive debris present. It seems rather to depend on other factors, such as the size of the haematom, degree of muscle necrosis, etc.

Для цитирования: Afifi K F, Jacob H A. [Wear measurements of hip prostheses with UHMW polyethylene (RCH-1000) socket and chromium plated protasul-10 head (author's transl)].. Zeitschrift fur Orthopadie und ihre Grenzgebiete. 1981. DOI: 10.1055/s-2008-1051438.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.