DOI 10.1007/s00117-024-01344-7

Медицинская библиотека · Radiologie (Heidelberg, Germany) · 2024

Detection of myeloma-associated osteolytic bone lesions with energy-integrating and photon-counting detector CT.

Grözinger Martin, Wennmann Markus, Sawall Stefan, Wehrse Eckhard, Sedaghat Sam, Neelsen Christian, Bauer Fabian, Goldschmidt Hartmut, Weru Vivienn, Ziener Christian H, Kopp-Schneider Annette, Schlemmer Heinz-Peter, Rotkopf Lukas T

СтатьяПеревод в очереди

Аннотация

BACKGROUND

A recent innovation in computed tomography (CT) imaging has been the introduction of photon-counting detector CT (PCD-CT) systems, which are able to register the number and the energy level of incoming x‑ray photons and have smaller detector elements compared with conventional CT scanners that operate with energy-integrating detectors (EID-CT).

OBJECTIVES

The study aimed to evaluate the potential benefits of a novel, non-CE certified PCD-CT in detecting myeloma-associated osteolytic bone lesions (OL) compared with a state-of-the-art EID-CT.

MATERIALS AND

METHODS

Nine patients with multiple myeloma stage III (according to Durie and Salmon) underwent magnetic resonance imaging (MRI), EID-CT, and PCD-CT of the lower lumbar spine and pelvis. The PCD-CT and EID-CT images of all myeloma lesions that were visible in clinical MRI scans were reviewed by three radiologists for corresponding OL. Additionally, the visualization of destructions to cancellous or cortical bone, and trabecular structures, was compared between PCD-CT and EID-CT.

RESULTS

Readers detected 21% more OL in PCD-CT than in EID-CT images (138 vs. 109; p < 0.0001). The sensitivity advantage of PCD-CT in lesion detection increased with decreasing lesion size. The visualization quality of cancellous and cortical destructions as well as of trabecular structures was rated higher by all three readers in PCD-CT images (mean image quality improvements for PCD-CT over EID-CT were +0.45 for cancellous and +0.13 for cortical destructions).

CONCLUSIONS

For myeloma-associated OL, PCD-CT demonstrated significantly higher sensitivity, especially with small size. Visualization of bone tissue and lesions was considered significantly better in PCD-CT than in EID-CT. This implies that PCD-CT scanners could potentially be used in the early detection of myeloma-associated bone lesions.

Для цитирования: Grözinger Martin, Wennmann Markus, Sawall Stefan, Wehrse Eckhard, Sedaghat Sam, Neelsen Christian, Bauer Fabian, Goldschmidt Hartmut, Weru Vivienn, Ziener Christian H, Kopp-Schneider Annette, Schlemmer Heinz-Peter, Rotkopf Lukas T. Detection of myeloma-associated osteolytic bone lesions with energy-integrating and photon-counting detector CT.. Radiologie (Heidelberg, Germany). 2024. DOI: 10.1007/s00117-024-01344-7.