DOI 10.1007/s00590-024-04114-2
Clinical evaluation of a multiplex PCR-based test for joint infection: a prospective diagnostic accuracy study of forty-nine patients.
Аннотация
PURPOSE: The purpose of this study was to evaluate the diagnostic accuracy (sensitivity, specificity, positive predictive value (PPV) and negative predictive value (NPV)) of the PCR-based BioFire® Joint Infection Panel (BJI Panel) against microbiological culture growth for patients suspected of having a native or prosthetic joint infection.
METHODS
Synovial fluid and tissue biopsies were prospectively collected from patients from June 2022 to June 2023. The results of the BJI Panel were compared with those of culture growth.
RESULTS
51 samples were included. Including all pathogens, the sensitivity was 69%, the specificity 89%, the PPV 73% and the NPV 86%. Including only pathogens in the BJI Panel, the sensitivity was 100%, the specificity 90%, the PPV 73% and the NPV 100%.
CONCLUSION
The BJI Panel has a high accuracy for detecting the pathogens in its panel, but the absence of important common pathogens from the panel reduces its sensitivity and NPV. With a short turnaround time and precise pathogen detection, the BJI Panel has the potential to add value as a complementary diagnostic method.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.