DOI 10.1097/bsd.0b013e318281608c
Rasterstereographic evaluation of interobserver and intraobserver reliability in postsurgical adolescent idiopathic scoliosis patients.
Аннотация
STUDY
DESIGN
Evaluation of rasterstereographic examinations in postsurgical adolescent scoliosis patients: a reliability study.
OBJECTIVE
To investigate the interobserver and intraobserver reliability of rasterstereographic 3-dimensional back surface analysis and reconstruction of spine parameters including trunk length, trunk inclination, lateral deviation, kyphotic and lordotic angles in adolescent idiopathic scoliosis patients after scoliosis surgery.
SUMMARY OF BACKGROUND DATA: The rasterstereography device formetric III 4D has been in routine clinical use for patients with scoliosis and other spinal deformities, reducing the patients' radiation exposure considerably. The reliability of this device has previously been examined in healthy volunteers but not in patients after scoliosis surgery.
MATERIALS AND
METHODS
Thirty-nine adolescent idiopathic scoliosis patients (32 women and 7 men) after scoliosis surgery with a mean age of 23.5 years (±10.6 y) were examined rasterstereographically by 5 investigators. Each investigator made a series of 3 measurements of each participating patient consecutively. The intraclass correlation coefficient (ICC) and the Pearson product moment correlation were calculated. In addition, the comparative analysis of the first and 15th measurements was assessed.
RESULTS
The ICCs of all investigated parameters demonstrated very high interobserver and intraobserver reliability. The maximum ICC (0.988) for interobserver reliability of the 5 investigators was found for the trunk length and the minimum ICC (0.918) for lateral deviation. There were highly significant positive correlations between the first, second, and third measurements (r=0.994-0.697; P<0.001) for intraobserver reliability. Overall highly significant correlation was found comparing the results of the first and the 15th rasterstereographic measurement.
CONCLUSIONS
The reliability of rasterstereography formetric III 4D in regard to the investigated parameters is excellent. The device can be used very efficiently by 1 or multiple investigators to analyze postsurgical adolescent idiopathic scoliosis patients in clinical practice, requiring only a single investigational exposure.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.