Design and biomechanical evaluation of a rodent spinal fixation device.

Shahrokni M, Zhu Q, Liu J, Tetzlaff W, Oxland T R

Оценочное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

STUDY

DESIGN

An in vitro and in vivo study in rats.

OBJECTIVES

To design a novel rat spinal fixation device and investigate its biomechanical effectiveness in stabilizing the spine up to 8 weeks post injury.

METHODS

A fixation device made of polyetheretherketone was designed to stabilize the spine via bilateral clamping pieces. The device effectiveness was assessed in a Sprague-Dawley rat model after it was applied to a spine with a fracture-dislocation injury produced at C5-C6. Animals were euthanized either immediately (n=6) or 8 weeks (n=9) post-injury and the C3-T1 segment of the cervical spine was removed for biomechanical evaluation. Segments of intact spinal columns (C3-T1) (n=6) served as uninjured controls. In these tests, anterior-posterior shear forces were applied to the C3 vertebra to produce flexion and extension bending moments at the injury site (peak 12.8 Nmm). The resultant two-dimensional motions at the injury site (that is, C5-C6) were measured using digital imaging and reported as ranges of motion (ROM) or neutral zones (NZ).

RESULTS

Flexion/extension ROMs (average±s.d.) were 18.1±3.3°, 19.9±7.5° and 1.5±0.7°, respectively for the intact, injured/fixed, and injured/8-week groups, with the differences being highly significant for the injured/8-week group (P=0.0002). Flexion/extension NZs were 3.4±2.8°, 5.0±2.4°, and 0.7±0.5°, respectively for the intact, injured/fixed, and injured/8-week groups, with the differences being significant for the injured/8-week group (P=0.04).

CONCLUSION

The device acutely stabilizes the spine and promotes fusion at the site of injury.

Для цитирования: Shahrokni M, Zhu Q, Liu J, Tetzlaff W, Oxland T R. Design and biomechanical evaluation of a rodent spinal fixation device.. Spinal cord. 2012. DOI: 10.1038/sc.2011.185.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.