Biomechanical analysis of compression screw fixation versus standard in situ pinning in slipped capital femoral epiphysis.
Аннотация
A slipped capital femoral epiphysis was created in 12 matched pairs of immature bovine femora using an anterior-to-posterior-directed shear force. All soft tissues, with the exception of the perichondrial ring, were removed before testing. One specimen from each pair was fixed with a single cannulated screw in standard fashion, whereas the contralateral specimen was fixed with a single screw that compressed the physis. The amount of compression achieved was quantified using Fuji film. Standard fixation yielded 1.4 MPa of pressure across the physis; compression fixation yielded 3.2 MPa, a 2.3-fold difference (p = 0.0001). The compression fixation was 47% more stiff than standard technique (p = 0.030), yet the differences in ultimate strength (p = 0.180) and energy absorbed at failure (p = 0.910) were not statistically significant. The stiffness of the compressed specimens remained less than that of the intact femora. Single-screw compression fixation of in vitro bovine femora was significantly more stiff than the current, widely used noncompression fixation technique, yet does not compromise the ultimate strength, energy absorbed, or the technical ease of single-implant fixation.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.