Clinical Application of an Augmented Reality Navigation System for Transforaminal Epidural Injection: A Randomized Controlled Trial.
Аннотация
Objectives
Augmented reality (AR) navigation systems are emerging to simplify and enhance the precision of medical procedures. Lumbosacral transforaminal epidural injection is a commonly performed procedure for the treatment and diagnosis of radiculopathy. Accurate needle placement while avoiding critical structures remains a challenge. For this purpose, we conducted a randomized controlled trial for our augmented reality navigation system.
Methods
This randomized controlled study involved 28 patients, split between a traditional C-arm guided group (control) and an AR navigation guided group (AR-NAVI), to compare procedure efficiency and radiation exposure. The AR-NAVI group used a real-time tracking system displaying spinal structure and needle position on an AR head-mounted display. The procedural time and C-arm usage (radiation exposure) were measured.
Results
All patients underwent successful procedures without complications. The AR-NAVI group demonstrated significantly reduced times and C-arm usage for needle entry to the target point (58.57 ± 33.31 vs. 124.91 ± 41.14, p < 0.001 and 3.79 ± 1.97 vs. 8.86 ± 3.94, p < 0.001).
Conclusions
The use of the AR navigation system significantly improved procedure efficiency and safety by reducing time and radiation exposure, suggesting a promising direction for future enhancements and validation.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.