Accuracy and digital screw path design of TiRobot-assisted pedicle screw placement for lumbar spondylolisthesis.
Аннотация
PURPOSE: To investigate lumbar spondylolisthesis screw placement assisted by TiRobot in terms of digital screw path design, accurate implementation, and accuracy evaluation method.
METHODS
In this study, we enrolled 40 patients with lumbar spondylolisthesis between December 2020 and August 2021 who underwent spine surgery at the Affiliated Hospital of PuTian University. Pre-operative computed tomography position and screw path designation, intra-operative pedicle screw placement according to pre-operative planning, and post-operative evaluation of the accuracy of screw placement were performed. 3D coordinates of the entry and exit points before and after the operation were collected. The qualified points at different levels of accuracy were counted. The screw placement accuracy was based on the absolute difference using the Chi-squared test.
RESULTS
In total, 194 screws were successfully implanted with no screws penetrating the cortex. The absolute difference of entry points X, Y, and Z coordinates before and after the operation was 0.425 ± 0.294 mm, 0.417 ± 0.310 mm, and 0.466 ± 0.327 mm, respectively. The corresponding values in terms of exit points were 0.702 ± 0.470 mm, 0.963 ± 0.595mm, and 0.983 ± 0.566 mm, respectively. No obvious differences in coordinates before and after the operation were observed with an entry point degree of accuracy of ≥ 1.2 mm and exit point degree of accuracy of ≥ 2.1 mm. Therefore, the real surgery was consistent with the design.
CONCLUSIONS
TiRobot-assisted lumbar spondylolisthesis surgery achieved optimal path designation and precise surgery.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.