Spinolaminar locking plates improve fixation strength compared to pedicle screws: a biomechanical analysis.

Kondrashov Dimitriy G, Piple Amit S, Ungurean Victor, Schlauch Adam M, Rowland Andrea, Tran Tien, Denisov Anton, Zaborovskii Nikita, Raji Oluwatodimu Richard

СтатьяСравнительное исследованиеПеревод в очереди

Аннотация

INTRODUCTION

Pedicle screw loosening is a significant complication of posterior spinal fixation, particularly among osteoporotic patients and in deformity constructs. In orthopedic trauma surgery, locking plates and screws have revolutionized the fixation of osteoporotic fractures. We have combined the traumatology principle of fixed-angle locking plate fixation with the spine principles of segmental instrumentation.

METHODS

A novel spinolaminar locking plate was designed based on morphometric studies of human thoracolumbar vertebrae. The plates were fixed to cadaveric human lumbar spines and connected to form 1-level L1-L2 or L4-L5 constructs and compared to similar pedicle screw constructs. Pure moment testing was performed to assess range of motion before and after 30,000 cycles of cyclic fatigue. Post-fatigue fixture pullout strength was assessed by applying a continuous axial tensile force oriented to the principal axis of the pedicle until pullout was observed.

RESULTS

Spinolaminar plate fixation resulted in superior pullout strength compared to pedicle screws (1,065 ± 400N vs. 714 ± 284N, p = 0.028). Spinolaminar plates performed equivalently to pedicle screws in range of motion reduction during flexion/extension and axial rotation. Pedicle screws outperformed the spinolaminar plates in lateral bending. Finally, no spinolaminar constructs failed during cyclic fatigue testing, whereas one pedicle screw construct did.

CONCLUSIONS

The spinolaminar locking plate maintained adequate fixation post-fatigue, particularly in flexion/extension and axial rotation compared to pedicle screws. Moreover, spinolaminar plates were superior to pedicle screw fixation with respect to cyclic fatiguing and pullout strength. The spinolaminar plates offer a viable option for posterior lumbar instrumentation in the adult spine.

Для цитирования: Kondrashov Dimitriy G, Piple Amit S, Ungurean Victor, Schlauch Adam M, Rowland Andrea, Tran Tien, Denisov Anton, Zaborovskii Nikita, Raji Oluwatodimu Richard. Spinolaminar locking plates improve fixation strength compared to pedicle screws: a biomechanical analysis.. Spine deformity. 2023. DOI: 10.1007/s43390-023-00716-8.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.