DOI 10.1097/bsd.0b013e31827a32ee

Медицинская библиотека · Journal of spinal disorders & techniques · 2015

3D-FSE Isotropic MRI of the Lumbar Spine: Novel Application of an Existing Technology.

Blizzard Daniel J, Haims Andrew H, Lischuk Andrew W, Arunakul Rattalerk, Hustedt Joshua W, Grauer Jonathan N

СтатьяПеревод в очереди

Аннотация

STUDY

DESIGN

Retrospective diagnostic trial.

OBJECTIVE

To determine the diagnostic performance of 3-dimensional isotropic fast/turbo spin-echo (3D-TSE) in routine lumbar spine magnetic resonance imaging (MRI).

SUMMARY OF BACKGROUND DATA: Conventional 2-dimensional fast spin-echo (2D-FSE) MRI requires independent acquisition of each desired imaging plane. This is time consuming and potentially problematic in spine imaging, as the plane of interest varies along the vertical axis due to lordosis, kyphosis, or possible deformity. 3D-TSE provides the capability to acquire volumetric data sets that can be dynamically reformatted to create images in any desired plane.

METHODS

Eighty subjects scheduled for routine lumbar MRI were included in a retrospective trial. Each subject underwent both 3D-TSE and conventional 2D-FSE axial and sagittal MRI sequences. For each subject, the 3D-TSE and 2D-FSE sequences were separately evaluated (minimum 4 wk apart) in a randomized order and read independently by 4 reviewers. Images were evaluated using specific criteria for stenosis, herniation, and degenerative changes.

RESULTS

The intermethod reliability for the 4 reviewers was 85.3%. Modified intermethod reliability analysis, disregarding disagreements between the lowest 2 descriptors for appropriate criteria (equivalent to "none" and "mild"), revealed average overall agreement of 94.6%. Using the above, modified criteria, interobserver variability for 3D-TSE was 89.1% and 88.3% for 2D-FSE (P=0.05), and intraobserver variability for 3D-TSE was 87.2% and 82.0% for 2D-FSE (P<0.01). The intermethod agreement between 3D-TSE and 2D-FSE was statistically noninferior to intraobserver 2D-FSE variability (P<0.01).

CONCLUSIONS

This systematic evaluation showed that there is a very high degree of agreement between diagnostic findings assessed on 3D-TSE and conventional 2D-FSE sequences. Overall, intermethod agreement was statistically noninferior to the intraobserver agreement between repeated 2D-FSE evaluations. Overall, this study shows that 3D-TSE performs equivalently, if not superiorly to 2D-FSE sequences. Reviewers found particular utility for the ability to manipulate image planes with the 3D-TSE if there was greater pathology or anatomic variation.

Для цитирования: Blizzard Daniel J, Haims Andrew H, Lischuk Andrew W, Arunakul Rattalerk, Hustedt Joshua W, Grauer Jonathan N. 3D-FSE Isotropic MRI of the Lumbar Spine: Novel Application of an Existing Technology.. Journal of spinal disorders & techniques. 2015. DOI: 10.1097/bsd.0b013e31827a32ee.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.