BACKGROUND

The efficacy of the powered rasp, a new reciprocating motion device for arthroscopic resection of osteophytes, has not been verified. The aim of this study was to compare the intraoperative efficacy of the powered rasp in arthroscopic resection of anterior ankle osteophytes to that of the conventional burr.

METHODS

A total of 49 consecutive patients who underwent arthroscopic resection of anterior ankle osteophytes (26 patients with the conventional burr and 23 patients with the powered rasp) were retrospectively reviewed. The preoperative volume of each osteophyte was measured using computerized tomography scan and three-dimensional software. The resection time was measured by review of the individual arthroscopy video, and the estimated resection rate was calculated as the volume of osteophytes/resection time.

RESULTS

The preoperative volume of osteophytes was not different between the two groups (847.8 ± 685.3 mm3 in the conventional burr and 913.3 ± 605.8 mm3 in the powered rasp, p = 0.726). The resection time was 442.4 ± 216.6 s (seconds) in the conventional burr and 386.4 ± 186.3 s in the powered rasp, and the estimated resection rate was 1.8 ± 1.0 mm3/s with the conventional burr and 2.4 ± 1.3 mm3/s with the powered rasp. These measurements were not significantly different between the two groups (p = 0.340 and 0.083, respectively).

CONCLUSION

The intraoperative efficacy of the powered rasp did not show superiority to that of the conventional burr in arthroscopic resection of anterior ankle osteophytes.

LEVEL OF EVIDENCE: Level III, retrospective comparative study.

Для цитирования: Park Young Hwan, Kim Woon, Choi Jung Woo, Choi Gi Won, Kim Hak Jun. Comparison of the intraoperative efficacy of the powered rasp and conventional burr in arthroscopic resection of anterior ankle osteophytes.. Foot and ankle surgery : official journal of the European Society of Foot and Ankle Surgeons. 2020. DOI: 10.1016/j.fas.2020.12.009.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.