PURPOSE: The primary purpose of this study was to retrospectively assess the incidence of bioabsorbable screw failure in skeletally immature patients treated for osteochondritis dissecans (OCD) of the knee. The second purpose was to assess implant degradation, focusing on differential breakdown of the intraosseous and extraosseous parts of the screw on magnetic resonance imaging (MRI).

METHODS

In this retrospective study, 24 patients (30 knees) with MRI-confirmed OCD were treated with a total of 61 biodegradable screws and followed up for a minimum of 2 years or until the onset of new symptoms. MRI scans were performed every 6 months to assess differences in signal intensity between the head and body of the screw. The angle between the surface of the head and the body of the screw was measured, with an angle greater than 90° interpreted as indicating a broken screw (i.e., failed implant).

RESULTS

Seven screws (all 2.7 mm, 11.5% of all screws) in 5 patients were considered broken screws. The implant failed completely in an additional 4 patients with breakage of 7 of 9 screws; 3 patients were considered late failures, occurring after more than 6 months. Altogether, 14 of a total of 61 screws (23%) were broken.

CONCLUSIONS

Screw breakage is a surprisingly frequent cause of failure in resorbable OCD fixation in skeletally immature patients. MRI data showed differential decomposition of the screw within and outside of bone as a possible cause.

LEVEL OF EVIDENCE: Level IV, therapeutic case series.

Для цитирования: Camathias Carlo, Gögüs Ulas, Hirschmann Michael T, Rutz Erich, Brunner Reinald, Haeni David, Vavken Patrick. Implant failure after biodegradable screw fixation in osteochondritis dissecans of the knee in skeletally immature patients.. Arthroscopy : the journal of arthroscopic & related surgery : official publication of the Arthroscopy Association of North America and the International Arthroscopy Association. 2014. DOI: 10.1016/j.arthro.2014.08.032.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.