Vertebral Endplate Concavity in Lateral Lumbar Interbody Fusion: Tapered 3D-Printed Porous Titanium Cage versus Squared PEEK Cage.

Segi Naoki, Nakashima Hiroaki, Shinjo Ryuichi, Kagami Yujiro, Machino Masaaki, Ito Sadayuki, Ouchida Jun, Morishita Kazuaki, Oishi Ryotaro, Yamauchi Ippei, Imagama Shiro

СтатьяПеревод в очереди

Аннотация

Background and

Objectives

To prevent postoperative problems in extreme lateral interbody fusion (XLIF), it is critical that the vertebral endplate not be injured. Unintentional endplate injuries may depend on the cage. A novel porous titanium cage for XLIF has improved geometry with a tapered tip and smooth surface. We hypothesized that this new cage should lead to fewer endplate injuries. Materials and

Methods

This retrospective study included 32 patients (mean 74.1 ± 6.7 years, 22 females) who underwent anterior and posterior combined surgery with XLIF for lumbar degenerative disease or adult spinal deformity from January 2018 to June 2022. A tapered 3D porous titanium cage (3DTi; 11 patients) and a squared PEEK cage (sPEEK; 21 patients) were used. Spinal alignment values were measured on X-ray images. Vertebral endplate concavity (VEC) was defined as concavity ≥ 1 mm of the endplate on computed tomography (CT) images, which were evaluated preoperatively and at 1 week and 3 months postoperatively.

Results

There were no significant differences in the patient demographic data and preoperative and 3-month postoperative spinal alignments between the groups. A 3DTi was used for 25 levels and an sPEEK was used for 38 levels. Preoperative local lordotic angles were 4.3° for 3DTi vs. 4.7° for sPEEK (p = 0.90), which were corrected to 12.3° and 9.1° (p = 0.029), respectively. At 3 months postoperatively, the angles were 11.6° for 3DTi and 8.2° for sPEEK (p = 0.013). VEC was present in 2 levels (8.0%) for 3DTi vs. 17 levels (45%) for sPEEK (p = 0.002). After 3 months postoperatively, none of the 3DTi had VEC progression; however, eight (21%) levels in sPEEK showed VEC progression (p = 0.019).

Conclusions

The novel 3DTi cage reduced endplate injuries by reducing the endplate load during cage insertion.

Для цитирования: Segi Naoki, Nakashima Hiroaki, Shinjo Ryuichi, Kagami Yujiro, Machino Masaaki, Ito Sadayuki, Ouchida Jun, Morishita Kazuaki, Oishi Ryotaro, Yamauchi Ippei, Imagama Shiro. Vertebral Endplate Concavity in Lateral Lumbar Interbody Fusion: Tapered 3D-Printed Porous Titanium Cage versus Squared PEEK Cage.. Medicina (Kaunas, Lithuania). 2023. DOI: 10.3390/medicina59020372.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.