MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1089/dia.2021.0319

Медицинская библиотека · Diabetes technology & therapeutics · 2021

Safety and Glycemic Outcomes During the MiniMed™ Advanced Hybrid Closed-Loop System Pivotal Trial in Adolescents and Adults with Type 1 Diabetes.

Carlson Anders L, Sherr Jennifer L, Shulman Dorothy I, Garg Satish K, Pop-Busui Rodica, Bode Bruce W, Lilenquist David R, Brazg Ron L, Kaiserman Kevin B, Kipnes Mark S, Thrasher James R, Reed John H Chip, Slover Robert H, Philis-Tsimikas Athena, Christiansen Mark, Grosman Benyamin, Roy Anirban, Vella Melissa, Jonkers Richard A M, Chen Xiaoxiao, Shin John, Cordero Toni L, Lee Scott W, Rhinehart Andrew S, Vigersky Robert A

СтатьяМногоцентровое исследованиеПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

Introduction

This trial assessed safety and effectiveness of an advanced hybrid closed-loop (AHCL) system with automated basal (Auto Basal) and automated bolus correction (Auto Correction) in adolescents and adults with type 1 diabetes (T1D). Materials and

Methods

This multicenter single-arm study involved an intent-to-treat population of 157 individuals (39 adolescents aged 14-21 years and 118 adults aged ≥22-75 years) with T1D. Study participants used the MiniMed™ AHCL system during a baseline run-in period in which sensor-augmented pump +/- predictive low glucose management or Auto Basal was enabled for ∼14 days. Thereafter, Auto Basal and Auto Correction were enabled for a study phase (∼90 days), with glucose target set to 100 or 120 mg/dL for ∼45 days, followed by the other target for ∼45 days. Study endpoints included safety events and change in mean A1C, time in range (TIR, 70-180 mg/dL) and time below range (TBR, <70 mg/dL). Run-in and study phase values were compared using Wilcoxon signed-rank test or paired t-test.

Results

Overall group time spent in closed loop averaged 94.9% ± 5.4% and involved only 1.2 ± 0.8 exits per week. Compared with run-in, AHCL reduced A1C from 7.5% ± 0.8% to 7.0% ± 0.5% (<0.001, Wilcoxon signed-rank test, n = 155), TIR increased from 68.8% ± 10.5% to 74.5% ± 6.9% (<0.001, Wilcoxon signed-rank test), and TBR reduced from 3.3% ± 2.9% to 2.3% ± 1.7% (<0.001, Wilcoxon signed-rank test). Similar benefits to glycemia were observed for each age group and were more pronounced for the nighttime (12 AM-6 AM). The 100 mg/dL target increased TIR to 75.4% (n = 155), which was further optimized at a lower active insulin time (AIT) setting (i.e., 2 h), without increasing TBR. There were no severe hypoglycemic or diabetic ketoacidosis events during the study phase.

Conclusions

These findings show that the MiniMed AHCL system is safe and allows for achievement of recommended glycemic targets in adolescents and adults with T1D. Adjustments in target and AIT settings may further optimize glycemia and improve user experience. Clinical Trial Registration number: NCT03959423.

Для цитирования: Carlson Anders L, Sherr Jennifer L, Shulman Dorothy I, Garg Satish K, Pop-Busui Rodica, Bode Bruce W, Lilenquist David R, Brazg Ron L, Kaiserman Kevin B, Kipnes Mark S, Thrasher James R, Reed John H Chip, Slover Robert H, Philis-Tsimikas Athena, Christiansen Mark, Grosman Benyamin, Roy Anirban, Vella Melissa, Jonkers Richard A M, Chen Xiaoxiao, Shin John, Cordero Toni L, Lee Scott W, Rhinehart Andrew S, Vigersky Robert A. Safety and Glycemic Outcomes During the MiniMed™ Advanced Hybrid Closed-Loop System Pivotal Trial in Adolescents and Adults with Type 1 Diabetes.. Diabetes technology & therapeutics. 2021. DOI: 10.1089/dia.2021.0319.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.