Safety and Glycemic Outcomes During the MiniMed™ Advanced Hybrid Closed-Loop System Pivotal Trial in Adolescents and Adults with Type 1 Diabetes.
Аннотация
Introduction
This trial assessed safety and effectiveness of an advanced hybrid closed-loop (AHCL) system with automated basal (Auto Basal) and automated bolus correction (Auto Correction) in adolescents and adults with type 1 diabetes (T1D). Materials and
Methods
This multicenter single-arm study involved an intent-to-treat population of 157 individuals (39 adolescents aged 14-21 years and 118 adults aged ≥22-75 years) with T1D. Study participants used the MiniMed™ AHCL system during a baseline run-in period in which sensor-augmented pump +/- predictive low glucose management or Auto Basal was enabled for ∼14 days. Thereafter, Auto Basal and Auto Correction were enabled for a study phase (∼90 days), with glucose target set to 100 or 120 mg/dL for ∼45 days, followed by the other target for ∼45 days. Study endpoints included safety events and change in mean A1C, time in range (TIR, 70-180 mg/dL) and time below range (TBR, <70 mg/dL). Run-in and study phase values were compared using Wilcoxon signed-rank test or paired t-test.
Results
Overall group time spent in closed loop averaged 94.9% ± 5.4% and involved only 1.2 ± 0.8 exits per week. Compared with run-in, AHCL reduced A1C from 7.5% ± 0.8% to 7.0% ± 0.5% (<0.001, Wilcoxon signed-rank test, n = 155), TIR increased from 68.8% ± 10.5% to 74.5% ± 6.9% (<0.001, Wilcoxon signed-rank test), and TBR reduced from 3.3% ± 2.9% to 2.3% ± 1.7% (<0.001, Wilcoxon signed-rank test). Similar benefits to glycemia were observed for each age group and were more pronounced for the nighttime (12 AM-6 AM). The 100 mg/dL target increased TIR to 75.4% (n = 155), which was further optimized at a lower active insulin time (AIT) setting (i.e., 2 h), without increasing TBR. There were no severe hypoglycemic or diabetic ketoacidosis events during the study phase.
Conclusions
These findings show that the MiniMed AHCL system is safe and allows for achievement of recommended glycemic targets in adolescents and adults with T1D. Adjustments in target and AIT settings may further optimize glycemia and improve user experience. Clinical Trial Registration number: NCT03959423.