DOI 10.3389/fsurg.2025.1722297
Novel anatomically contoured locking plate shows improved anatomic conformity vs. a conventional T-shaped plate in anterior popliteus transtibial-tuberosity high tibial osteotomy: a radiologic comparison.
Аннотация
BACKGROUND
High tibial osteotomy (HTO) corrects varus malalignment and unloads the medial knee compartment, yet the conventional T-shaped locking plate may cause discomfort due to suboptimal anatomic conformity. We developed a novel locking plate to improve anatomic fit and fixation stability and compared it with the T-shaped plate in Anterior Popliteus Transtibial-tuberosity high tibial osteotomy (APTT-HTO).
METHODS
This retrospective comparative study included 14 consecutive patients who underwent APTT-HTO between August 2024 and June 2025. Patients were grouped by implant type (novel plate: n = 7; T-shaped plate: n = 7). Postoperative CT quantified plate orientation/position, plate-bone conformity (gap and position mismatch), screw alignment, and standard alignment parameters [hip-knee-ankle angle [HKA], posterior tibial slope [PTS]].
RESULTS
The novel plate showed more posteromedial placement and superior anatomic conformity. The plate angle relative to the posterior tibial condylar reference line was larger with the novel plate (74.49° ± 8.76°) than with the T-shaped plate (62.62° ± 7.05°, P < 0.05). The proximal central screw-plate angle was smaller (5.70° ± 4.80° vs. 27.48° ± 6.05°, P < 0.05), the plate-bone gap was reduced (1.47 (1.35-2.28) vs. 3.12 (2.70-3.26) mm, P < 0.05), and plate position mismatch was lower (18.86 (14.32-23.87) vs. 34.31 (30.57-42.08) %, P < 0.05). Proximal and distal offsets considered separately were not significantly different (P > 0.05). Both groups achieved the planned coronal correction (P < 0.05), and the sagittal PTS remained unchanged (P < 0.05).
CONCLUSIONS
In this CT-based cohort, the novel anatomically contoured locking plate achieved superior anatomic conformity in APTT-HTO-characterized by more posteromedial positioning, a more favorable screw trajectory, and improved plate-bone apposition-compared with a conventional T-shaped plate. These findings warrant confirmation in larger, prospective studies.