MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1177/10711007221136539

Медицинская библиотека · Foot & ankle international · 2022

Early-Term Results of the Cadence Total Ankle Prosthesis: An European Noninventor Study.

Malherbe Corentin, Deleu Paul-André, Devos Bevernage Bernhard, Birch Ivan, Maldague Pierre, Gombault Vincent, Putzeys Pit, Leemrijse Thibaut

СтатьяКомментарийПеревод в очереди

Аннотация

BACKGROUND

Although considerable literature can be found on the outcome of total ankle replacement (TAR), only a few studies have reported the results of the fixed-bearing Cadence prosthesis. This noninventor study reports a consecutive series of 60 Cadence TAR systems with a mean of 2.9 years' follow-up, focusing on clinical and radiographic outcomes and early complications. This study is the first to assess true postoperative radiographic ankle prosthesis range of motion (ROM) and to report an unanticipated serious adverse device effect.

METHODS

Sixty patients who underwent primary TAR with the Cadence prosthesis between July 2016 and July 2019 were clinically and radiographically evaluated preoperatively and at last follow-up after the procedure. Revisions, additional procedures, implant failure, and complications were reported according to the classifications of Vander Griend and Glazebrook. Radiographic outcomes included radiographic TAR ROM, bone-implant interface, and alignment parameters.

RESULTS

The survival rate of the prosthesis was 98.3%. The mean radiographic ankle ROM at the last follow-up was 24 degrees (9 degrees of dorsiflexion and 15 degrees of plantarflexion). The coronal and sagittal alignment of TAR was 90.8 degrees and 3.9 degrees, respectively. Bone-implant interface analysis revealed osteolysis in 9 ankles (15%) and radiolucent lines in 33 ankles (55%) occurring at both component interfaces. Intraoperative complications were 3 periprosthetic malleolar fractures (5%). Five talar implant fractures (implant failure of 8.3%) were observed, and 1 unexplained persistent pain that required a conversion from TAR to a tibiotalocalcaneal arthrodesis.

CONCLUSION

Clinical, radiograph ROM, implant position outcomes, and survival rate at an early-term follow-up of 2.9 years were similar to those reported in recent Cadence studies. However, this study reports 5 unanticipated talar implant fractures and a high rate of posterior radiolucent lines.

LEVEL OF EVIDENCE: Level IV, retrospective case series.

Для цитирования: Malherbe Corentin, Deleu Paul-André, Devos Bevernage Bernhard, Birch Ivan, Maldague Pierre, Gombault Vincent, Putzeys Pit, Leemrijse Thibaut. Early-Term Results of the Cadence Total Ankle Prosthesis: An European Noninventor Study.. Foot & ankle international. 2022. DOI: 10.1177/10711007221136539.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.