Effect of vitamin E-enhanced highly cross-linked polyethylene on wear rate and particle debris in anatomic total shoulder arthroplasty: a biomechanical comparison to ultrahigh-molecular-weight polyethylene.

Khan Adam Z, Maxwell Michael J, Parrott Russell M, Bowler Adam R, Glass Evan A, Miller Drew, Vasconcellos Daniel, Brownhill James R, Austin Luke S, Cuff Derek J, Murthi Anand M, Smith Matthew J, Wiater J Michael, Jawa Andrew

СтатьяСравнительное исследованиеПеревод в очереди

Аннотация

BACKGROUND

Particle-induced osteolysis resulting from polyethylene wear remains a source of implant failure in anatomic total shoulder designs. Modern polyethylene components are irradiated in an oxygen-free environment to induce cross-linking, but reducing the resulting free radicals with melting or heat annealing can compromise the component's mechanical properties. Vitamin E has been introduced as an adjuvant to thermal treatments. Anatomic shoulder arthroplasty models with a ceramic head component have demonstrated that vitamin E-enhanced polyethylene show improved wear compared with highly cross-linked polyethylene (HXLPE). This study aimed to assess the biomechanical wear properties and particle size characteristics of a novel vitamin E-enhanced highly cross-linked polyethylene (VEXPE) glenoid compared to a conventional ultrahigh-molecular-weight polyethylene (UHMWPE) glenoid against a cobalt chromium molybdenum (CoCrMo) head component.

METHODS

Biomechanical wear testing was performed to compare the VEXPE glenoid to UHMWPE glenoid with regard to pristine polyethylene wear and abrasive endurance against a polished CoCrMo alloy humeral head in an anatomic shoulder wear-simulation model. Cumulative mass loss (milligrams) was recorded, and wear rate calculated (milligrams per megacycle [Mc]). Under pristine wear conditions, particle analysis was performed, and functional biologic activity (FBA) was calculated to estimate particle debris osteolytic potential. In addition, 95% confidence intervals for all testing conditions were calculated.

RESULTS

The average pristine wear rate was statistically significantly lower for the VEXPE glenoid compared with the HXLPE glenoid (0.81 ± 0.64 mg/Mc vs. 7.00 ± 0.45 mg/Mc) (P < .05). Under abrasive wear conditions, the VEXPE glenoid had a statistically significant lower average wear rate compared with the UHMWPE glenoid comparator device (18.93 ± 5.80 mg/Mc vs. 40.47 ± 2.63 mg/Mc) (P < .05). The VEXPE glenoid demonstrated a statistically significant improvement in FBA compared with the HXLPE glenoid (0.21 ± 0.21 vs. 1.54 ± 0.49 (P < .05).

CONCLUSIONS

A new anatomic glenoid component with VEXPE demonstrated significantly improved pristine and abrasive wear properties with lower osteolytic particle debris potential compared with a conventional UHMWPE glenoid component. Vitamin E-enhanced polyethylene shows early promise in shoulder arthroplasty components. Long-term clinical and radiographic investigation needs to be performed to verify if these biomechanical wear properties translate to diminished long-term wear, osteolysis, and loosening.

Для цитирования: Khan Adam Z, Maxwell Michael J, Parrott Russell M, Bowler Adam R, Glass Evan A, Miller Drew, Vasconcellos Daniel, Brownhill James R, Austin Luke S, Cuff Derek J, Murthi Anand M, Smith Matthew J, Wiater J Michael, Jawa Andrew. Effect of vitamin E-enhanced highly cross-linked polyethylene on wear rate and particle debris in anatomic total shoulder arthroplasty: a biomechanical comparison to ultrahigh-molecular-weight polyethylene.. Journal of shoulder and elbow surgery. 2024. DOI: 10.1016/j.jse.2023.11.016.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.