Empirical planning target volume modeling for high precision MRI guided intracranial radiotherapy.

Stewart James, Sahgal Arjun, Zadeh Mahtab M, Moazen Bahareh, Jabehdar Maralani Pejman, Breen Stephen, Lau Angus, Binda Shawn, Keller Brian, Husain Zain, Myrehaug Sten, Detsky Jay, Soliman Hany, Tseng Chia-Lin, Ruschin Mark

СтатьяПеревод в очереди

Аннотация

PURPOSE: Magnetic resonance image-guided radiotherapy for intracranial indications is a promising advance; however, uncertainties remain for both target localization after translation-only MR setup and intrafraction motion. This investigation quantified these uncertainties and developed a population-based planning target volume (PTV) model to explore target and organ-at-risk (OAR) volumetric coverage tradeoffs.

METHODS

Sixty-six patients, 49 with a primary brain tumor and 17 with a post-surgical resection cavity, treated on a 1.5T-based MR-linac across 1329 fractions were included. At each fraction, patients were setup by translation-only fusion of the online T1 MRI to the planning image. Each fusion was independently repeated offline accounting for rotations. The six degree-of-freedom difference between fusions was applied to transform the planning CTV at each fraction (CTVfx). A PTV model parameterized by volumetric CTVfx coverage, proportion of fractions, and proportion of patients was developed. Intrafraction motion was quantified in a 412 fraction subset as the fusion difference between post- and pre-irradiation T1 MRIs.

RESULTS

For the left-right/anterior-posterior/superior-inferior axes, mean ± SD of the rotational fusion differences were 0.1 ± 0.8/0.1 ± 0.8/-0.2 ± 0.9°. Covering 98 % of the CTVfx in 95 % of fractions in 95 % of patients required a 3 mm PTV margin. Margin reduction decreased PTV-OAR overlap; for example, the proportion of optic chiasm overlapped by the PTV was reduced up to 23.5 % by margin reduction from 4 mm to 3 mm.

CONCLUSIONS

An evidence-based PTV model was developed for brain cancer patients treated on the MR-linac. Informed by this model, we have clinically adopted a 3 mm PTV margin for conventionally fractionated intracranial patients.

Для цитирования: Stewart James, Sahgal Arjun, Zadeh Mahtab M, Moazen Bahareh, Jabehdar Maralani Pejman, Breen Stephen, Lau Angus, Binda Shawn, Keller Brian, Husain Zain, Myrehaug Sten, Detsky Jay, Soliman Hany, Tseng Chia-Lin, Ruschin Mark. Empirical planning target volume modeling for high precision MRI guided intracranial radiotherapy.. Clinical and translational radiation oncology. 2023. DOI: 10.1016/j.ctro.2023.100582.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.