PURPOSE: The incidence of atlanto-axial injuries is continuously increasing and often requires surgical treatment. Recently, Harati developed a new procedure combining polyaxial transarticular screws with polyaxial atlas massae lateralis screws via a rod system with promising clinical results, yet biomechanical data is lacking. This biomechanical study consequently aims to evaluate the properties of the Harati technique.

METHODS

Two groups, each consisting of 7 cervical vertebral segments (C1/2), were formed and provided with a dens axis type 2 fracture according to Alonzo. One group was treated with the Harms and the other with the Harati technique. The specimen was loaded via a lever arm to simulate extension, flexion, lateral flexion and rotation. For statistical analysis, dislocation (°) was measured and compared.

RESULTS

For extension and flexion, the Harati technique displayed a mean dislocation of 4.12° ± 2.36° and the Harms technique of 8.48° ± 1.49° (p < 0.01). For lateral flexion, the dislocation was 0.57° ± 0.30° for the Harati and 1.19° ± 0.25° for the Harms group (p < 0.01). The mean dislocation for rotation was 1.09° ± 0.48° for the Harati and 2.10° ± 0.31° for the Harms group (p < 0.01). No implant failure occurred.

CONCLUSION

This study found a significant increase in biomechanical stability of the Harati technique when compared to the technique by Harms et al. Consequently, this novel technique can be regarded as a promising alternative for the treatment of atlanto-axial instabilities.

Для цитирования: Cavalcanti Kußmaul Adrian, Kühlein Titus, Greiner Axel, Walter Sandy, Becker Christopher A, Kistler Manuel, Rubenbauer Bianka, Andreß Sebastian, Böcker Wolfgang, Bruder Jan. Improving stability of atlantoaxial fusion: a biomechanical study.. European journal of orthopaedic surgery & traumatology : orthopedie traumatologie. 2022. DOI: 10.1007/s00590-022-03465-y.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.