Biomechanical evaluation and comparison of polyetheretherketone rod system to traditional titanium rod fixation.

Ponnappan Ravi K, Serhan Hassan, Zarda Brett, Patel Ravi, Albert Todd, Vaccaro Alexander R

Сравнительное исследованиеСтатьяПеревод в очереди

Аннотация

BACKGROUND CONTEXT: Polyetheretherketone (PEEK) has been increasingly used as a biomaterial for spinal implants. PEEK lumbar fusion rods have recently become available for use in posterior lumbar fusion procedures.

PURPOSE: To compare Polyetheretherketone Rod System to traditional titanium rod fixation in a cadaveric model and provide mechanical test data for the PEEK system.

STUDY

DESIGN

Biomechanical testing.

METHODS

Cadaveric biomechanical testing was conducted to compare Expedium 5.5 mm PEEK rods to titanium rods of equivalent diameter. Biomaterials testing was performed to determine static and dynamic performance of Expedium 5.5 mm PEEK rods with 6% BaSo4 in compressive bending and torsion.

RESULTS

Cadaveric testing demonstrated that PEEK rods can significantly reduce the range of motion of a destabilized segment. The testing showed no significant difference in the stability provided by PEEK and titanium rods in posterolateral fusion (PLF) or posterior lumbar interbody fusion (PLIF) constructs. PEEK static compressive bending tests showed 67 degrees displacement without fracture of the rod. Torsion testing showed 30 degrees of rotation without yield or plastic deformation. Dynamic compression testing revealed two fatigue runouts at 23 degrees.

CONCLUSIONS

PEEK rods provide comparable stability to titanium rods of equivalent diameter in cadaveric testing. Mechanical testing suggests PEEK rods can withstand far beyond the angular displacements suggested by cadaveric testing and that of normal physiologic range of motion. Potential advantages to PEEK rods include better anterior column load sharing, reduced stress at bone-to-screw interface, and reduced computed tomography and magnetic resonance imaging scatter and artifact.

Для цитирования: Ponnappan Ravi K, Serhan Hassan, Zarda Brett, Patel Ravi, Albert Todd, Vaccaro Alexander R. Biomechanical evaluation and comparison of polyetheretherketone rod system to traditional titanium rod fixation.. The spine journal : official journal of the North American Spine Society. 2008. DOI: 10.1016/j.spinee.2008.08.002.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.