Медицинская библиотека · The spine journal : official journal of the North American Spine Society · 2008

X-Stop: foot drop.

Epstein Nancy E

Клинический случайСтатьяПеревод в очереди

Аннотация

BACKGROUND CONTEXT: The X-Stop (Kyphon, Sunnyvale, CA), a device implanted between adjacent lumbar spinous processes at one or two levels particularly in the geriatric population, theoretically decompresses the spinal canal by reversing the lordosis.

PURPOSE: The X-Stop relieves the symptoms of lumbar stenosis.

STUDY DESIGN/

SETTING

A case report from the United States.

PATIENT SAMPLE: A case report focusing on how X-Stop placement resulted in a bilateral foot drop in an 84-year-old patient, which was relieved 9 months later with laminectomy.

OUTCOME MEASURES: The patient's outcome was based on the neurological examination.

METHODS

An 84-year-old male with lumbar neurogenic claudication/radiculopathy, exhibited magnetic resonance (MR)-documented L4-L5 lumbar stenosis and Grade I degenerative spondylolisthesis. At an outside institution, an L4-L5 X-Stop was placed; he immediately developed a bilateral foot drop. Three months later, the X-Stop extruded, and was removed; his foot drop remained unchanged. Nine months after the original surgery, based on new MR/computed tomography (CT) documented severe (L2-L3, L4-L5), moderate (L1-L2, L3-L4, L5-S1) stenosis, and the L4-L5 Grade I degenerative spondylolisthesis, the second surgeon performed an L1-S1 laminectomy with L4-L5 noninstrumented fusion.

RESULTS

Postoperatively, the patient's bilateral foot drop largely resolved.

CONCLUSIONS

Placing the X-Stop device at severely stenotic levels, particularly when accompanied by degenerative spondylolisthesis, may contribute to significant postoperative neurological sequelae in geriatric patients. For those with limited comorbidities, operative decompression may prove the safer alternative.

Для цитирования: Epstein Nancy E. X-Stop: foot drop.. The spine journal : official journal of the North American Spine Society. 2008. DOI: 10.1016/j.spinee.2008.08.004.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.