Are There Thermal Risks Associated With Dry Arthroscopy? A Simulated Comparison of Temperatures in Wet Versus Dry Wrist Arthroscopy.

Namazie Ridzwan, Huynh Minh Nq, Nedunchezhian Ishvar, Bindra Randipsingh

СтатьяСравнительное исследованиеПеревод в очереди

Аннотация

PURPOSE: Dry arthroscopy of the wrist is comparable in safety to wet arthroscopy and is preferred in some cases as it avoids fluid extravasation. However, there is a potential for increased temperatures within the joint and risk of tissue damage. The purpose of this study was to document the maximum temperatures measured in a wrist simulator under dry and wet conditions using different arthroscopes and light sources.

METHODS

A custom designed testing rig with an enclosed space was fabricated with ports to allow for insertion of a temperature probe, arthroscope, and drainage port. Three different arthroscope sizes (2.7, 2.4, and 1.9 mm) and two different light sources (Xenon and light-emitting diode) were tested in wet and dry conditions.

RESULTS

Three tested combinations caused the recorded temperature to exceed the temperature threshold. All were using the Xenon light source and with direct contact with the tip of the arthroscope. Light intensity was 100% for the 1.9 mm arthroscope, 60% for the 2.4 mm arthroscope, and 80% for the 2.7 mm arthroscope. Maximum temperatures reached for the arthroscopes were 45.1°C, 55°C, and 50.7°C, respectively. The times to exceed the safe temperature were 620, 380, and 490 seconds, respectively.

CONCLUSIONS

Generally, the temperatures measured with dry arthroscopy are safe. Larger diameter scopes and Xenon light sources generate more heat.

CLINICAL RELEVANCE: We propose that surgeons should use a light-emitting diode light source at the minimum intensity required, avoid direct contact with tissues and during heat generating procedures, such as burring, and intermittently flush the joint with saline to dissipate heat.

Для цитирования: Namazie Ridzwan, Huynh Minh Nq, Nedunchezhian Ishvar, Bindra Randipsingh. Are There Thermal Risks Associated With Dry Arthroscopy? A Simulated Comparison of Temperatures in Wet Versus Dry Wrist Arthroscopy.. The Journal of hand surgery. 2026. DOI: 10.1016/j.jhsa.2026.01.026.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.