Медицинская библиотека · Archives of orthopaedic and trauma surgery · 2022

A new knee arthrometer demonstrated to be reliable and accurate to assess anterior tibial translation in comparison with stress radiographs.

Milano Giuseppe, Colosio Alessandro, Scaini Alessandra, Motta Marcello, Raggi Andrea, Zanoni Fabio, Galli Stefano, Saccomanno Maristella F

СтатьяПеревод в очереди

Аннотация

INTRODUCTION

For several years, many arthrometers have been developed to assess anterior knee laxity. The aim of our study was to evaluate the validity of a new practical and handy testing device with the hypothesis that the new arthrometer had good validity in terms of reliability and accuracy.

METHODS

Lachman test was performed on five fresh frozen cadaveric knees by five examiners. Anterior tibial translation (ATT) was measured with a new arthrometer (BLU-DAT) and on lateral stress radiographs. Data on ATT were obtained under 7 kg (69 N), 9 kg (88 N), and maximum manual traction (MMT). Tests were performed on the same specimens before and after arthroscopic ACL excision. Inter-rater reliability of the BLU-DAT measures was assessed with the intraclass correlation coefficient (ICC) for single and average measurements. The Bland-Altman method was used to estimate agreement between the BLU-DAT and stress radiographs.

RESULTS

ICC values for single measurements were 0.62, 0.54 and 0.58 for 7-kg, 9-kg and MMT assessment, respectively. Overall reliability was good (ICC = 0.63). ICC values for average measurements were 0.89, 0.85 and 0.88 for 7-kg, 9-kg and MMT assessment, respectively. Overall reliability was very good (ICC = 0.90) SEM ranged from 1.4 mm to 1.6 mm for single measurements and was below 1 mm at each testing condition for average measurements. Analysis of agreement between BLU-DAT and radiographic measurements showed a mean difference equal to 0.83 mm ± 2.1 mm (95% CI: 0.55-1.11). Upper LOA was equal to 4.9 mm (95% CI: 5.39-4.41). Lower LOA was equal to - 3.2 mm (95% CI: - 2.71 to - 3.69).

CONCLUSION

Measurement of anterior knee laxity with the BLU-DAT testing device has a good to very good inter-rater reliability and good agreement with a gold standard such as stress radiographs. Cadaveric Diagnostic Study, Level of Evidence IV.

Для цитирования: Milano Giuseppe, Colosio Alessandro, Scaini Alessandra, Motta Marcello, Raggi Andrea, Zanoni Fabio, Galli Stefano, Saccomanno Maristella F. A new knee arthrometer demonstrated to be reliable and accurate to assess anterior tibial translation in comparison with stress radiographs.. Archives of orthopaedic and trauma surgery. 2022. DOI: 10.1007/s00402-022-04679-9.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.