DOI 10.1007/s00402-022-04679-9
A new knee arthrometer demonstrated to be reliable and accurate to assess anterior tibial translation in comparison with stress radiographs.
Аннотация
INTRODUCTION
For several years, many arthrometers have been developed to assess anterior knee laxity. The aim of our study was to evaluate the validity of a new practical and handy testing device with the hypothesis that the new arthrometer had good validity in terms of reliability and accuracy.
METHODS
Lachman test was performed on five fresh frozen cadaveric knees by five examiners. Anterior tibial translation (ATT) was measured with a new arthrometer (BLU-DAT) and on lateral stress radiographs. Data on ATT were obtained under 7 kg (69 N), 9 kg (88 N), and maximum manual traction (MMT). Tests were performed on the same specimens before and after arthroscopic ACL excision. Inter-rater reliability of the BLU-DAT measures was assessed with the intraclass correlation coefficient (ICC) for single and average measurements. The Bland-Altman method was used to estimate agreement between the BLU-DAT and stress radiographs.
RESULTS
ICC values for single measurements were 0.62, 0.54 and 0.58 for 7-kg, 9-kg and MMT assessment, respectively. Overall reliability was good (ICC = 0.63). ICC values for average measurements were 0.89, 0.85 and 0.88 for 7-kg, 9-kg and MMT assessment, respectively. Overall reliability was very good (ICC = 0.90) SEM ranged from 1.4 mm to 1.6 mm for single measurements and was below 1 mm at each testing condition for average measurements. Analysis of agreement between BLU-DAT and radiographic measurements showed a mean difference equal to 0.83 mm ± 2.1 mm (95% CI: 0.55-1.11). Upper LOA was equal to 4.9 mm (95% CI: 5.39-4.41). Lower LOA was equal to - 3.2 mm (95% CI: - 2.71 to - 3.69).
CONCLUSION
Measurement of anterior knee laxity with the BLU-DAT testing device has a good to very good inter-rater reliability and good agreement with a gold standard such as stress radiographs. Cadaveric Diagnostic Study, Level of Evidence IV.