MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1097/bot.0000000000002552

Медицинская библиотека · Journal of orthopaedic trauma · 2023

Biomechanical Comparison of Intramedullary Versus Extramedullary Implants for Fixation of Simple Pertrochanteric Fractures.

Schader Jana F, Zderic Ivan, Dauwe Jan, Sommer Christoph, Gueorguiev Boyko, Stoffel Karl

СтатьяПеревод в очереди

Аннотация

OBJECTIVES

To evaluate the biomechanical performance of the intramedullary TFN-ADVANCED Proximal Femoral Nailing System (TFNA) versus the extramedullary Femoral Neck System (FNS) for fixation of simple pertrochanteric fractures in a human cadaveric model.

METHODS

Ten human cadaveric femoral pairs were implanted pairwise with either TFNA or FNS. A simple pertrochanteric fracture OTA/AO 31-A1 was created and all specimens were biomechanically tested under progressively increasing cyclic loading until failure. Interfragmentary and bone-implant movements were monitored by motion tracking.

RESULTS

Axial stiffness was comparable between TFNA and FNS, P = 0.34. Similarly, varus deformation, femoral head rotation around neck axis and implant migration remained without significant differences between TFNA and FNS after 3000 cycles (800 N peak load), P ≥ 0.10. However, cycles to 15-mm leg shortening were significantly higher for TFNA versus FNS, P < 0.01.

CONCLUSIONS

From a biomechanical perspective, with its current design, FNS does not seem to be a valid alternative to TFNA for treatment of simple pertrochanteric fractures.

Для цитирования: Schader Jana F, Zderic Ivan, Dauwe Jan, Sommer Christoph, Gueorguiev Boyko, Stoffel Karl. Biomechanical Comparison of Intramedullary Versus Extramedullary Implants for Fixation of Simple Pertrochanteric Fractures.. Journal of orthopaedic trauma. 2023. DOI: 10.1097/bot.0000000000002552.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.