Cervical pedicle screws vs. lateral mass screws: uniplanar fatigue analysis and residual pullout strengths.

Johnston Todd L, Karaikovic Eldin E, Lautenschlager Eugene P, Marcu David

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND CONTEXT: Although successful clinical use of cervical pedicle screws has been reported, anatomical studies have shown the possibility for serious iatrogenic injury. However, there are only a limited number of reports on the biomechanical properties of these screws which evaluate the potential benefits of their application.

PURPOSE: To investigate if the pull-out strengths after cyclic uniplanar loading of cervical pedicle screws are superior to lateral mass screws.

STUDY

DESIGN

An in vitro biomechanical study.

METHODS

Twenty fresh-frozen disarticulated human vertebrae (C3-C7) were randomized to receive both a 3.5 mm cervical pedicle screw and lateral mass screw. The screws were cyclically loaded 200 times in the sagittal plane. The amount of displacement was recorded every 50 cycles. After cyclical loading, the screws were pulled and tensile load to failure was recorded. Bone density was measured in each specimen and maximum screw insertion torque was recorded for each screw.

RESULTS

During loading the two screw types showed similar stability initially, however the lateral mass screws rapidly loosened compared to the pedicle screws. The rate of loosening in the lateral mass screws was widely variable, while the performance of the pedicle screws was very consistent. The pullout strengths were significantly higher for the cervical pedicle screws (1214 N vs. 332 N) and 40% failed by fracture of the pedicle rather than screw pullout. Pedicle screw pullout strengths correlated with both screw insertion torque and specimen bone density.

CONCLUSIONS

Cervical pedicle screws demonstrated a significantly lower rate of loosening at the bone-screw interface, as well as higher strength after fatigue testing. These biomechanical strengths may justify their use in certain limited clinical applications.

Для цитирования: Johnston Todd L, Karaikovic Eldin E, Lautenschlager Eugene P, Marcu David. Cervical pedicle screws vs. lateral mass screws: uniplanar fatigue analysis and residual pullout strengths.. The spine journal : official journal of the North American Spine Society. 2006. DOI: 10.1016/j.spinee.2006.03.019.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.