DOI 10.1097/ftd.0000000000000310
Comparison of Three Assays to Quantify Infliximab, Adalimumab, and Etanercept Serum Concentrations.
Аннотация
BACKGROUND
To optimize treatment of inflammatory diseases, interest in the measurement of anti-tumor necrosis factor alpha (anti-TNFα) serum drug concentrations is increasing. Preferably, assays for the detection of these drugs should be compared using the same reference material. In this study, 2 commercially available enzyme-linked immunosorbent assays (ELISAs) and a commercially available bioassay for the determination of anti-TNFα drugs are compared.
METHODS
Serum samples from infliximab-, adalimumab-, and etanercept-treated patients, control samples from ustekinumab-treated patients, and healthy donors were obtained. ELISAs manufactured by Sanquin and Theradiag and the iLite reporter gene-based bioassay from Biomonitor were compared.
RESULTS
Sanquin, Theradiag, and iLite assays concordantly (100%) detected infliximab, adalimumab, and etanercept in the relevant patient groups. The Sanquin ELISAs specifically detected the anti-TNFα drug they were designed for, whereas the Theradiag and iLite showed cross-reactivity with other anti-TNFα drugs. Ustekinumab was not detected in any of the assays. Sanquin, Theradiag, and iLite exhibited linear quantitative correlation for all drug concentration assays. However, there were statistically significant quantitative differences in measured concentrations.
CONCLUSIONS
All 3 commercially available assays seem suitable for therapeutic drug monitoring of anti-TNFα drugs, allowing sensitive and comparable detection of infliximab, adalimumab, and etanercept concentrations, however with differences in specificity and recovery.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.