MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.wneu.2022.05.056

Медицинская библиотека · World neurosurgery · 2022

Comparison of the Susceptibility to Implant Failure in the Lateral, Posterior, and Transforaminal Lumbar Interbody Fusion: A Finite Element Analysis.

Oikawa Ryo, Murakami Hideki, Endo Hirooki, Yan Hirotaka, Yamabe Daisuke, Chiba Yusuke, Oikawa Ryosuke, Nishida Norihiro, Chen Xian, Sakai Takashi, Doita Minoru

СтатьяПеревод в очереди

Аннотация

OBJECTIVE

There are several techniques for lumbar interbody fusion, and implant failure following lumbar interbody fusion can be troublesome. This study aimed to compare the stress in posterior implant and peri-screw vertebral bodies among lateral lumbar interbody fusion (LLIF), posterior lumbar interbody fusion (PLIF), and transforaminal lumbar interbody fusion (TLIF) and to select the technique that is least likely to cause implant failure.

METHODS

We created an intact L3-L5 model and simulated the LLIF, PLIF, and TLIF techniques at L4-L5 using finite element methods. All models at the lower portion of L5 were fixed and imposed a preload of 400 N and a moment of 7.5 Nm on the upper portion of L3 to simulate flexion, extension, lateral bending, and axial rotation. We investigated the peak stresses and stress concentration in the posterior implant and peri-screw vertebral bodies for the LLIF, PLIF, and TLIF techniques.

RESULTS

The extension, flexion, bending, and rotation peak stresses and stress concentration in the posterior implant, as well as the peri-screw vertebral bodies, were the lowest in LLIF, followed by PLIF and TLIF.

CONCLUSIONS

It was found that implant failure was least likely to occur in LLIF, followed by PLIF and TLIF. Hence, surgeons should be aware of these factors when selecting an appropriate surgical technique and be careful for implant failure during postoperative follow-up.

Для цитирования: Oikawa Ryo, Murakami Hideki, Endo Hirooki, Yan Hirotaka, Yamabe Daisuke, Chiba Yusuke, Oikawa Ryosuke, Nishida Norihiro, Chen Xian, Sakai Takashi, Doita Minoru. Comparison of the Susceptibility to Implant Failure in the Lateral, Posterior, and Transforaminal Lumbar Interbody Fusion: A Finite Element Analysis.. World neurosurgery. 2022. DOI: 10.1016/j.wneu.2022.05.056.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.