MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1053/apmr.2002.32439

Медицинская библиотека · Archives of physical medicine and rehabilitation · 2002

Radiation exposure of the spinal interventionalist performing fluoroscopically guided lumbar transforaminal epidural steroid injections.

Botwin Kenneth P, Thomas Santhosh, Gruber Robert D, Torres Francisco M, Bouchlas Constantine C, Rittenberg Joshua J, Rao Sanjiv

СтатьяПеревод в очереди

Аннотация

OBJECTIVE

To evaluate radiation exposure to spinal interventionalists while performing transforaminal epidural steroid injections (TFESIs).

DESIGN

Prospective study.

SETTING

Multidisciplinary spine center.

PARTICIPANTS

One hundred consecutive patients with either herniated nucleus pulposus (HNP) or lumbar spinal stenosis (LSS).

INTERVENTION: Fluroscopically guided lumbar TFESIs.

MAIN OUTCOME MEASURE: Radiation exposure was monitored by radiography technologists who allocated 4 dosimetry badges to all spinal interventionalists performing fluroscopically guided lumbar TFESIs on patients being treated for radicular pain. Badges were placed on the ring finger, glasses, and the inside and outside of the lead apron worn by the interventionalists. The radiography technologists also wore marked badges outside their lead aprons. One control badge was placed 67in away from the fluoroscopy table and a second badge was placed in a desk more than 500ft away from the procedure to monitor ambient radiation.

RESULTS

The average fluoroscopy time per procedure was 15.16 seconds. The average exposure per procedure was 0.7mrem at the ring badge, 0.4mrem at the glasses badge, and 0.3mrem at the outside apron badge. No radiation was detectable at the inside apron or at the outside room control badge. The cumulative exposure to the interventionalists from all 100 procedures was 70mrem at the ring badge, 40mrem at the glasses badge, and 30mrem at the outside apron badge. The radiography technologists' average exposure during these procedures was below the limit of detectablility. Radiation time under fluoroscopy ranged from 5 to 38 seconds. The interventionalist's exposure to radiation was significantly greater during procedures conducted on patients with LSS then during procedures on patients with HNP.

CONCLUSION

Adhering to a radiation safety program that includes maximizing the distance the spinal interventionalist is from the radiation source, decreasing exposure time, and proper shielding is essential when performing fluoroscopically guided lumbar TFESIs. Our study shows that exposure to radiation of the spinal interventionalist performing fluoroscopically guided lumbar TFESIs was well within safety limits when proper techniques were followed.

Для цитирования: Botwin Kenneth P, Thomas Santhosh, Gruber Robert D, Torres Francisco M, Bouchlas Constantine C, Rittenberg Joshua J, Rao Sanjiv. Radiation exposure of the spinal interventionalist performing fluoroscopically guided lumbar transforaminal epidural steroid injections.. Archives of physical medicine and rehabilitation. 2002. DOI: 10.1053/apmr.2002.32439.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.