MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.jhsa.2017.08.031

Медицинская библиотека · The Journal of hand surgery · 2017

The Rate of Radial Head Prosthesis Removal or Revision: A Systematic Review and Meta-Analysis.

Kachooei Amir R, Baradaran Aslan, Ebrahimzadeh Mohammad H, van Dijk C Niek, Chen Neal

СтатьяМетаанализСистематический обзорПеревод в очереди

Аннотация

PURPOSE: We conducted a meta-analysis and systematic review with the primary objective to determine the overall incidence of radial head prosthesis removal or revision. Our secondary objectives addressed the incidence of removal or revision based on the type of prosthesis fixation (cemented, uncemented smooth stem, uncemented press-fit), material (metal, Vitallium, titanium, pyrocarbon), and design (short vs long stem and monopolar vs bipolar), and the reasons for prosthetic removal or revision.

METHODS

We included 30 studies with a total of 1,017 patients out of whom 77 prostheses were removed and 45 prostheses were revised.

RESULTS

The pooled rate of radial head prosthesis removal or revision was 10.0% (95% confidence interval, 7.3%-13.6%) with a mean follow-up of 38 months. Subgroup analysis showed that the incidence of removal/revision was lowest with the cemented fixation, longer-stem, Vitallium material, and bipolar prosthesis. More than half of the prostheses were removed/revised for excision of the heterotopic ossification (47%) and for the treatment of stiffness and limitation of motion (42%). Other reasons recorded were pain (19%), loosening (16%), overstuffing (13%), instability (12%), infection (8%), and prosthesis disassembly (4%).

CONCLUSIONS

The current data show that the highest incidence of removal/revision occurred within 2 years after implantation. There was no major difference in the incidence of removal/revision among different designs and materials. Implant removal was often performed as part of a procedure to manage elbow stiffness and heterotopic ossification at the surgeon's preference, not necessarily because the implant was malfunctioning. It appears that most radial head arthroplasties have an acceptable and comparable mid-term longevity; however, it is unclear whether long-term longevity will differ between devices.

TYPE OF STUDY/LEVEL OF EVIDENCE: Prognostic II.

Для цитирования: Kachooei Amir R, Baradaran Aslan, Ebrahimzadeh Mohammad H, van Dijk C Niek, Chen Neal. The Rate of Radial Head Prosthesis Removal or Revision: A Systematic Review and Meta-Analysis.. The Journal of hand surgery. 2017. DOI: 10.1016/j.jhsa.2017.08.031.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.