How much different is the semi-quantification of synovitis according to the ultrasound system and the blood flow detection technology?

Iagnocco Annamaria, Martínez-Estupiñán Lina, Figus Fabiana, Olivas-Vergara Otto, Zabotti Alen, Borges Pablo, Agnes Cecilia, Zanetti Anna, Rozza Davide, Naredo Esperanza

СтатьяСравнительное исследованиеМногоцентровое исследованиеПеревод в очереди

Аннотация

AIM: To compare synovial blood flow scoring between different technologies and ultrasound (US) systems in active and inactive rheumatoid arthritis (RA).

MATERIAL AND

METHODS

Fifty-nine RA patients underwent B-mode, power Doppler (PD), colour Doppler (CD), B-Flow and High-Resolution (High-Res) PDI assessments of 6 joints with two US systems at two European centres. Each joint was semi-quantitatively scored for all ultrasound parameters. PD, CD and High-Res PDI synovial signal was also quantitatively scored.

RESULTS

Correlations between the total score of SH with system 1 and 2 were very high (≥ 0.90, p<0.0001). Baseline correlations between systems for PD and CD total scores were moderate to very high (0.44-0.96, p<0.05). At baseline, there were no significant differences between ultrasound systems for PD or CD semiquantitative-based total scores in active or inactive patients (p>0,05). B-Flow and High-Res total scores were significantly lower than PD or CD total scores (p<0.05).

CONCLUSION

A high-end and an entry-level US system were interchangeable for scoring SH and showed similar sensitivity and responsiveness in scoring synovial blood flow by PD and CD but not interchangeability. B-Flow and High-Res PDI were responsive, but they showed different sensitivity to detect synovial blood flow compared to conventional Doppler.

Для цитирования: Iagnocco Annamaria, Martínez-Estupiñán Lina, Figus Fabiana, Olivas-Vergara Otto, Zabotti Alen, Borges Pablo, Agnes Cecilia, Zanetti Anna, Rozza Davide, Naredo Esperanza. How much different is the semi-quantification of synovitis according to the ultrasound system and the blood flow detection technology?. Medical ultrasonography. 2024. DOI: 10.11152/mu-4391.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.