How much different is the semi-quantification of synovitis according to the ultrasound system and the blood flow detection technology?
Аннотация
AIM: To compare synovial blood flow scoring between different technologies and ultrasound (US) systems in active and inactive rheumatoid arthritis (RA).
MATERIAL AND
METHODS
Fifty-nine RA patients underwent B-mode, power Doppler (PD), colour Doppler (CD), B-Flow and High-Resolution (High-Res) PDI assessments of 6 joints with two US systems at two European centres. Each joint was semi-quantitatively scored for all ultrasound parameters. PD, CD and High-Res PDI synovial signal was also quantitatively scored.
RESULTS
Correlations between the total score of SH with system 1 and 2 were very high (≥ 0.90, p<0.0001). Baseline correlations between systems for PD and CD total scores were moderate to very high (0.44-0.96, p<0.05). At baseline, there were no significant differences between ultrasound systems for PD or CD semiquantitative-based total scores in active or inactive patients (p>0,05). B-Flow and High-Res total scores were significantly lower than PD or CD total scores (p<0.05).
CONCLUSION
A high-end and an entry-level US system were interchangeable for scoring SH and showed similar sensitivity and responsiveness in scoring synovial blood flow by PD and CD but not interchangeability. B-Flow and High-Res PDI were responsive, but they showed different sensitivity to detect synovial blood flow compared to conventional Doppler.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.