DOI 10.3233/thc-213248

Медицинская библиотека · Technology and health care : official journal of the European Society for Engineering and Medicine · 2022

Periprosthetic joint infection: Comparison of automated multiplex-PCR Unyvero i60 ITI cartridge system with bacterial culture and real-time PCR.

Lüdemann Martin, Sulastyanto Saskia, Raab Peter, Schoen Christoph, Rudert Maximilian

СтатьяПеревод в очереди

Аннотация

BACKGROUND

In the past, various efforts have been made to investigate diagnostic tools for periprosthetic-joint-infection (PJI). It is little-known about the diagnostic utility of polymerase-chain-reaction (PCR) in this context, especially concerning the role of multiplex-PCR assays comparing with conventional tissue culture.

OBJECTIVE

Evaluation of an automated-multiplex-PCR cartridge system for patients with suspicion of PJI in comparison with conventional microbiological culture and 16S-rDNA-PCR.

METHODS

On suspicion of PJI synovial fluid specimen were taken preoperatively or periprosthetic tissue was collected intraoperatively. Microbiological analysis included conventional culture, 16S-rDNA-PCR and automated-multiplex-PCR (Unyvero-i60-ITI®). The European-Bone-and-Joint-Infection-Society (EBJIS) criteria were used for PJI diagnosis. Positive and negative percent agreement was calculated. Total percentage agreement and Cohen's kappa coefficient were calculated. Sensitivity, specificity and positive predictive value of conventional culture, 16S-rDNA-PCR and multiplex-PCR were calculated. Ten specimens of proved PJI were used as control group.

RESULTS

Fifty specimen were suitable for culture. 14 (28%) were classified as PJI, 36 (72%) were aseptic. Coagulase-negative staphylococci was the most frequent detected pathogen. Concordance-rate between mPCR and culture results was 75.6% with a Cohen's kappa of 0.28. Concordance-rate between mPCR and 16S-rDNA was 82.9%, Cohen's kappa was 0.13. Concordance analysis between culture results and 16S-rDNA lead to a concordance-rate of 88.9%. Cohen's kappa was calculated with 0.6. With regard to the microbiological culture as reference, sensitivity of the mPCR was 0.33 and specificity was 0.91. Sensitivity and specificity of the 16S-rDNA-PCR was 0.55 and 0.97. The positive predictive value was 0.57 for the mPCR and 0.83 for the 16S-rDNA-PCR.

CONCLUSIONS

Due to fair agreement between mPCR and conventional microbiological culture, the tested multiplex-PCR could be an additional instrument for the detection of PJI but is not superior over the conventional culture.

Для цитирования: Lüdemann Martin, Sulastyanto Saskia, Raab Peter, Schoen Christoph, Rudert Maximilian. Periprosthetic joint infection: Comparison of automated multiplex-PCR Unyvero i60 ITI cartridge system with bacterial culture and real-time PCR.. Technology and health care : official journal of the European Society for Engineering and Medicine. 2022. DOI: 10.3233/thc-213248.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.