DOI 10.3233/bme-151561

Медицинская библиотека · Bio-medical materials and engineering · 2015

High sensitivity, high surface area Enzyme-linked Immunosorbent Assay (ELISA).

Singh Harpal, Morita Takahiro, Suzuki Yuma, Shimojima Masayuki, Le Van An, Sugamata Masami, Yang Ming

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

Enzyme-linked immunosorbent assays (ELISA) are considered the gold standard in the demonstration of various immunological reactions with an application in the detection of infectious diseases such as during outbreaks or in patient care.

OBJECTIVE

This study aimed to produce an ELISA-based diagnostic with an increased sensitivity of detection compared to the standard 96-well method in the immunologic diagnosis of infectious diseases.

METHODS

A '3DStack' was developed using readily available, low cost fabrication technologies namely nanoimprinting and press stamping with an increased surface area of 4 to 6 times more compared to 96-well plates. This was achieved by stacking multiple nanoimprinted polymer sheets. The flow of analytes between the sheets was enhanced by rotating the 3DStack and confirmed by Finite-Element (FE) simulation. An Immunoglobulin G (IgG) ELISA for the detection of antibodies in human serum raised against Rubella virus was performed for validation.

RESULTS

An improved sensitivity of up to 1.9 folds higher was observed using the 3DStack compared to the standard method.

CONCLUSIONS

The increased surface area of the 3DStack developed using nanoimprinting and press stamping technologies, and the flow pattern between sheets generated by rotating the 3DStack were potential contributors to a more sensitive ELISA-based diagnostic device.

Для цитирования: Singh Harpal, Morita Takahiro, Suzuki Yuma, Shimojima Masayuki, Le Van An, Sugamata Masami, Yang Ming. High sensitivity, high surface area Enzyme-linked Immunosorbent Assay (ELISA).. Bio-medical materials and engineering. 2015. DOI: 10.3233/bme-151561.