Magnetic Resonance Imaging Findings Following Ceramic-on-Ceramic Total Hip Arthroplasty Using Modular Neck Femoral Components.
Аннотация
BACKGROUND
Adverse reactions to metal debris after total hip arthroplasty are an increasing concern. Metal debris is released from the metal-on-metal articulation itself as well as from the head-neck taper. There is also a risk of corrosion at the neck-stem junction. Some stems were recalled because of high revision rates as a result of metal debris from the modular junction. We conducted a retrospective analysis of the 3-year to 8-year clinical and radiographic outcomes of a consecutive cohort of patients who underwent cementless total hip arthroplasty using a modular hip prosthesis with ceramic-on-ceramic bearings.
METHODS
We retrospectively evaluated 104 patients who had undergone 119 primary ceramic-on-ceramic total hip arthroplasties using modular neck femoral components between December 2006 and November 2012. All hips were implanted with a titanium alloy stem and titanium alloy neck. Clinical outcomes were evaluated using the Japanese Orthopaedic Association hip score. Prosthesis survival was calculated according to the Kaplan-Meier method considering any reason for revision as failure. Postoperative magnetic resonance imaging was performed and graded using a previously described pseudotumor grading system. Mean follow-up was 5.9 years (range: 3 to 8 years).
RESULTS
The mean age of the patients at the time of the index arthroplasty was 59.0 years. Mean hip score improved from 45.9 (range: 9 to 90) preoperatively to 91.5 (range: 65 to 100) at the final follow-up. The overall implant survival rate was 97.3% at 8 years when revision for any reason was used as the endpoint. Abnormalities on magnetic resonance imaging were seen in 16 of 91 hips (17.6%). All abnormalities were cystic thin-walled lesions located in the peritrochanteric area. There was no revision operation due to adverse reactions to metal debris.
CONCLUSIONS
A modular neck femoral prosthesis with ceramic-on-ceramic bearings offers satisfactory 3-year to 8-year results. However, abnormalities on magnetic resonance imaging were seen in one-fifth of the patients, although the clinical significance is unknown.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.