A biomechanical analysis of double-screw, double-button, and screw-button fixation constructs in patient-specific instrument-guided Latarjet procedure.

Ziegenfuss Brandon L, Launay Marine M, Maharaj Jashint C, Green Nicholas R, Stalin Kathir Azhagan, Barlow Gregory, Erbulut Deniz U, Whitehouse Sarah L, Vijaysegaran Praveen, Pivonka Peter, Cutbush Kenneth, Gupta Ashish

СтатьяПеревод в очереди

Аннотация

BACKGROUND

The Latarjet coracoid transfer procedure reliably stabilizes the glenohumeral joint for shoulder instability. However, complications such as graft osteolysis, nonunion and fracture continue to affect patient clinical outcomes. The double-screw (SS) construct is regarded as the gold standard method of fixation. SS constructs are associated with graft osteolysis. More recently, a double-button technique (BB) has been suggested to minimize graft-related complications. However, BB constructs are associated with fibrous nonunion. To mitigate this risk, a single screw combined with a single button (SB) construct has been proposed. It is thought that this technique incorporates the strength of the SS construct and allows micromotion superiorly to mitigate stress shielding-related graft osteolysis.

AIMS: The primary aim of this study was to compare the failure load of SS, BB, and SB constructs under a standardized biomechanical loading protocol. The secondary aim was to characterize the displacement of each construct throughout testing.

METHODS

Computed tomography scans of 20 matched-pair cadaveric scapulae were performed. Specimens were harvested and dissected free of soft tissue. SS and BB techniques were randomly assigned to specimens for matched-pair comparison with SB trials. A patient-specific instrument (PSI)-guided Latarjet procedure was performed on each scapula. Specimens were tested using a uniaxial mechanical testing device under cyclic loading (100 cycles, 1 Hz, 200 N/s) followed by a load-to-failure protocol (0.5 mm/s). Construct failure was defined by graft fracture, screw avulsion, and/or graft displacement of more than 5 mm.

RESULTS

Forty scapulae from 20 fresh frozen cadavers with a mean age of 69.3 years underwent testing. On average, SS constructs failed at 537.8 N (SD 296.8), whereas BB constructs failed at 135.1 N (SD 71.4). SB constructs required a significantly greater load to fail compared with BB constructs (283.5 N, SD 162.8, P = .039). Additionally, SS (1.9 mm, IQR 0.87) had a significantly lower maximum total graft displacement during the cyclic loading protocol compared with SB (3.8 mm, IQR 2.4, P = .007) and BB (7.4 mm, IQR 3.1, P < .001) constructs.

CONCLUSION

These findings support the potential of the SB fixation technique as a viable alternative to SS and BB constructs. Clinically, the SB technique could reduce the incidence of loading-related graft complications seen in the first 3 months of BB Latarjet cases. The study is limited to time-specific results and does not account for bone union or osteolysis.

Для цитирования: Ziegenfuss Brandon L, Launay Marine M, Maharaj Jashint C, Green Nicholas R, Stalin Kathir Azhagan, Barlow Gregory, Erbulut Deniz U, Whitehouse Sarah L, Vijaysegaran Praveen, Pivonka Peter, Cutbush Kenneth, Gupta Ashish. A biomechanical analysis of double-screw, double-button, and screw-button fixation constructs in patient-specific instrument-guided Latarjet procedure.. Journal of shoulder and elbow surgery. 2023. DOI: 10.1016/j.jse.2023.02.117.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.