DOI 10.1007/s00590-016-1889-1

Медицинская библиотека · European journal of orthopaedic surgery & traumatology : orthopedie traumatologie · 2016

Cadaveric assessment of a 3D-printed aiming device for implantation of a hinged elbow external fixator.

Dos Santos Alexandre, Creze Maud, Begin Matthieu, Laemmel Elisabeth, Assabah Bouchra, Soubeyrand Marc

Сравнительное исследованиеСтатьяПеревод в очереди

Аннотация

INTRODUCTION

Proper implantation of a hinged external elbow fixator (HEEF) is demanding since it requires precise alignment between the flexion-extension's and HEEF's axis. In order to optimize this alignment, we have developed a 3D-printed aiming device. The primary goal of the study was to compare the aiming device-based technique with the conventional pin technique. The secondary goal was to determine whether it is possible to share the aiming device with the surgical community.

MATERIALS AND

METHODS

A HEEF was implanted in cadavers with either the aiming device (n = 6) or the conventional pin technique (n = 6). For both techniques the duration of the procedure, the radiation exposure as well as the offset and angular divergence between the HEEF's and flexion-extension's axis were compared. To achieve the secondary goal, two surgeons used aiming devices 3D-printed from files sent by email in order to implant HEEF on cadaveric specimens (n = 6).

RESULTS

Duration of the procedure was not significantly different between both techniques. However, the aiming device allowed for reduction of the number of image intensifier shots (p = 0.005), angular divergence (p = 0.02) and offset between both axes (p = 0.05). The aiming devices have been delivered less than 15 days after ordering, and they have allowed proper implantation of six HEEF.

CONCLUSION

The 3D-printed aiming device allowed less irradiant and more accurate implantation of HEEF. It is possible to share it with other surgeons.

Для цитирования: Dos Santos Alexandre, Creze Maud, Begin Matthieu, Laemmel Elisabeth, Assabah Bouchra, Soubeyrand Marc. Cadaveric assessment of a 3D-printed aiming device for implantation of a hinged elbow external fixator.. European journal of orthopaedic surgery & traumatology : orthopedie traumatologie. 2016. DOI: 10.1007/s00590-016-1889-1.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.