MRgFUS ablation of a recurrent tenosynovial giant cell tumor in the foot using ExAblate 2100 system in combination with patient immobilization device.

Long Z, Hesley G K, Lu A, Hangiandreou N J, Gorny K R, Tiegs-Heiden C A

СтатьяКлинический случайПеревод в очереди

Аннотация

INTRODUCTION

Magnetic Resonance-guided Focused Ultrasound (MRgFUS) treatment for certain anatomy locations can be extremely challenging due to patient positioning and potential motion. This present study describes the treatment of a recurrent tenosynovial giant cell tumor of the plantar forefoot using the ExAblate 2100 system in combination with patient immobilization device.

METHODS

Prior to the treatment, several patient immobilization devices were investigated. Vacuum cushions were selected and tested for safety and compatibility with the treatment task and the MR environment.

RESULTS

During the treatment, one vacuum cushion immobilized the patient's right leg in knee flexion and allowed the bottom of the foot to be securely positioned on the treatment window. Another vacuum cushion supported the patient upper body extended outside the scanner bore. 19 sonications were successfully executed. The treatment was judged to be successful. No immediate complications were observed.

CONCLUSIONS

MRgFUS treatment of a recurrent tenosynovial giant cell tumor of the right plantar forefoot was successful with the use of patient immobilization vacuum cushions.

IMPLICATIONS FOR PRACTICE: The immobilization system could be utilized to aid future MRgFUS treatment of lesions in challenging anatomic locations. Various sizes of the vacuum cushions are available to potentially better accommodate other body parts and treatment configurations.

Для цитирования: Long Z, Hesley G K, Lu A, Hangiandreou N J, Gorny K R, Tiegs-Heiden C A. MRgFUS ablation of a recurrent tenosynovial giant cell tumor in the foot using ExAblate 2100 system in combination with patient immobilization device.. Radiography (London, England : 1995). 2024. DOI: 10.1016/j.radi.2024.03.010.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.