MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.ejrad.2022.110414

Медицинская библиотека · European journal of radiology · 2022

Bone visualization of the cervical spine with deep learning-based synthetic CT compared to conventional CT: A single-center noninferiority study on image quality.

van der Kolk Brigitta Britt Y M, Slotman Derk J Jorik, Nijholt Ingrid M, van Osch Jochen A C, Snoeijink Tess J, Podlogar Martin, van Hasselt Boudewijn A A M, Boelhouwers Henk J, van Stralen Marijn, Seevinck Peter R, Schep Niels W L, Maas Mario, Boomsma Martijn F

Сравнительное исследованиеEquivalence TrialСтатьяПеревод в очереди

Аннотация

PURPOSE: To investigate whether the image quality of a specific deep learning-based synthetic CT (sCT) of the cervical spine is noninferior to conventional CT.

METHOD: Paired MRI and CT data were collected from 25 consecutive participants (≥ 50 years) with cervical radiculopathy. The MRI exam included a T1-weighted multiple gradient echo sequence for sCT reconstruction. For qualitative image assessment, four structures at two vertebral levels were evaluated on sCT and compared with CT by three assessors using a four-point scale (range 1-4). The noninferiority margin was set at 0.5 point on this scale. Additionally, acceptable image quality was defined as a score of 3-4 in ≥ 80% of the scans. Quantitative assessment included geometrical analysis and voxelwise comparisons.

RESULTS

Qualitative image assessment showed that sCT was noninferior to CT for overall bone image quality, artifacts, imaging of intervertebral joints and neural foramina at levels C3-C4 and C6-C7, and cortical delineation at C6-C7. Noninferiority was weak to absent for cortical delineation at level C3-C4 and trabecular bone at both levels. Acceptable image quality was achieved for all structures in sCT and CT, except for trabecular bone in sCT and level C6-C7 in CT. Geometrical analysis of the sCT showed good to excellent agreement with CT. Voxelwise comparisons showed a mean absolute error of 80.05 (±6.12) HU, dice similarity coefficient (cortical bone) of 0.84 (±0.04) and structural similarity index of 0.86 (±0.02).

CONCLUSIONS

This deep learning-based sCT was noninferior to conventional CT for the general visualization of bony structures of the cervical spine, artifacts, and most detailed structure assessments.

Для цитирования: van der Kolk Brigitta Britt Y M, Slotman Derk J Jorik, Nijholt Ingrid M, van Osch Jochen A C, Snoeijink Tess J, Podlogar Martin, van Hasselt Boudewijn A A M, Boelhouwers Henk J, van Stralen Marijn, Seevinck Peter R, Schep Niels W L, Maas Mario, Boomsma Martijn F. Bone visualization of the cervical spine with deep learning-based synthetic CT compared to conventional CT: A single-center noninferiority study on image quality.. European journal of radiology. 2022. DOI: 10.1016/j.ejrad.2022.110414.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.