MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.jcms.2025.07.014

Медицинская библиотека · Journal of cranio-maxillo-facial surgery : official publication of the European Association for Cranio-Maxillo-Facial Surgery · 2025

Fitting accuracy of preformed osteosynthesis plates for zygomaticomaxillary complex fractures.

Raghoebar I I, Schreurs R, Youssef S A L Y, Helmers R, de Lange J, Dubois L

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Preformed osteosynthesis plates are a novel treatment option for the fixation of zygomaticomaxillary complex (ZMC) fractures. A preformed plate may improve the reduction accuracy, as an accurate fit and sufficient fixation possibilities are provided.

PURPOSE: This study aimed to evaluate the virtual fitting accuracy of preformed anatomical osteosynthesis zygoma plates and to assess whether their shape and size are adequate for ZMC fracture treatment.

METHODS

The fitting accuracy of three preformed osteosynthesis plates (KLS Martin: zygoma small, zygoma large, and zygoma-paranasal) was evaluated in 100 patients with unilateral ZMC fractures. Using a semi-automatic method, the plates were virtually positioned in their optimal fitting position on the mirrored 3D model of the unaffected side. Overall virtual adequacy was assessed by calculating the plate-to-bone distance and evaluating the fixation possibilities.

RESULTS

The median (IQR) plate-to-bone distances were 0.22 [0.16-0.28] mm (zygoma small), 0.25 [0.20-0.34] mm (zygoma large) and 0.51 [0.41-0.75] mm (zygoma-paranasal), demonstrating a significant improvement in fitting accuracy as the plate size decreased (p < 0.001). The median (IQR) number of available fixation points differed significantly: 6 [5-7] (zygoma small), 7 [6-8] (zygoma large), and 14 [12.25-16] (zygoma-paranasal) (p < 0.001). The zygoma small, zygoma large, and zygoma-paranasal plates were deemed adequate in 97 %, 98 %, and 93 % of the fractures, respectively.

CONCLUSIONS

Preformed anatomical osteosynthesis plates for ZMC reconstruction demonstrate an accurate fit, sufficient fixation possibilities, and overall virtual adequacy. These plates are designed to have a high level of anatomical congruence with the zygomatic bone, offering potential advantages in fracture reduction. However, successful clinical implementation requires careful attention to virtual surgical planning, and consideration of the plate rigidity and compulsive fit.

Для цитирования: Raghoebar I I, Schreurs R, Youssef S A L Y, Helmers R, de Lange J, Dubois L. Fitting accuracy of preformed osteosynthesis plates for zygomaticomaxillary complex fractures.. Journal of cranio-maxillo-facial surgery : official publication of the European Association for Cranio-Maxillo-Facial Surgery. 2025. DOI: 10.1016/j.jcms.2025.07.014.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.