OBJECTIVES

The Femoral Neck System (FNS) is a fixed-angle side plate device approved for use in fixation of femoral neck fractures. The FNS perforates the lateral cortex of the subtrochanteric femur, which may increase the risk of postoperative subtrochanteric fractures compared with the inverted triangle cannulated screw (CS) construct. The purpose of this biomechanical study was to compare forces required to create subtrochanteric fractures in FNS and CS constructs in a synthetic bone biomechanical model.

METHODS

Osteopenic sawbones were divided into five groups: Noninstrumented femurs, CS with the inferior screw proximal to the midpoint of the lesser trochanter ("normal position," CSN), CS with the inferior screw distal to the lesser trochanter ("low position," CSL), one-hole FNS, and two-hole FNS. Models were loaded to failure on a Servohydraulic Test System. Means and standard deviations were calculated for load to failure. Each group was paired and analyzed using two-sample t-tests.

RESULTS

Constructs in order of greatest to lowest load to failure include the following: Control (2950 N), one-hole FNS (2542 N), two-hole FNS (2444 N), CSN (2318 N), and CSL (2102 N). Load to failure was greater for the CSN group compared with CSL ( P = 0.04). Load to failure was greater for one-hole FNS versus CSL ( P = 0.002) and CSN ( P = 0.05). Load to failure was greater for two-hole FNS versus CSL ( P = 0.006) but not versus CSN ( P = 0.16). Load to failure was similar between one- and two-hole FNS ( P = 0.24).

CONCLUSION

The FNS demonstrated failure at the subtrochanteric femur in this synthetic femora biomechanical model. The load to failure of FNS was superior compared with CS placed below the level of the lesser trochanter. The load to failure of the one-hole FNS was also superior to CS placed proximal to the midpoint of the lesser trochanter.

Для цитирования: Kuttner Nicolas P, Cancio-Bello Alexandra M, Thompson Austen L, Fitzsimmons James S, Berglund Lawrence J, Sems Stephen A, Cross William W, Hidden Krystin A, Yuan Brandon J. Biomechanical Evaluation of the Risk of Subtrochanteric Fracture After Femoral Neck Fixation in a Synthetic Model: The Femoral Neck System versus Cannulated Screws.. The Journal of the American Academy of Orthopaedic Surgeons. 2025. DOI: 10.5435/jaaos-d-25-00469.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.