Initial Experience of Appendicular Fracture Fixation with Low Profile Titanium Alloy Locking Plates: A Retrospective Study of 30 Fractures.
Аннотация
OBJECTIVE
This retrospective study assessed the complications and mid-term outcomes of osteosynthesis using the String of Pearls Low Profile (SOP-LP) plate system in cats and small dogs with appendicular fractures.
METHODS
Clinical data from April 2018 to June 2021, involving 30 fractures in 29 cases, were reviewed. Fractures were stabilized using either the 1.5-mm or 2.0-mm SOP-LP plate. Data included patient signalment, fracture description, repair method, intraoperative complications and short-term follow-up outcomes. Midterm follow-up was performed by telephone-based veterinary questionnaires.
RESULTS
Of the 30 fractures, 3 had complications linked to the SOP-LP system. Overall, 7/30 had complications. Major complications (n = 6) included two implant failures, one implant-associated irritation and three Kirschner wire migrations. Implant failures (n = 2) were with the 1.5-mm plates stabilizing radius/ulna fractures in dogs weighing over 2.9 kg. Short-term radiographic follow-up demonstrated clinical or radiographic union in all cases. In all cases available (n = 18) for midterm follow-up, the function was found to be full or acceptable.
CONCLUSION
The SOP-LP plate can be used successfully for a wide variety of fractures in cats and small dogs; however, careful case selection is required, particularly when using the 1.5-mm plate as a single implant in patients weighing over 3 kg.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.